Hanging Product to Weigh
After the operator touches the check button to verify the kit has been
sealed and the red cell storage containers have been removed, the
touch screen will display the hang product prompt.
.
Typical Procedure Information Screen
To weigh the products, perform the following steps:
1.
Remove the plasma, red cell, and in-process containers from
their locations on the hanging scales.
2.
Hang one red cell storage container on the in-process scale.
3.
Hang the other red cell storage container on the plasma scale.
If necessary, open the clamps on the red cell storage
containers and allow the red cell product to equilibrate, so
there are approximately even amounts of product in each
storage container.
Note: If there is a large volume difference (> 10 mL) between the
two red cell storage containers, as indicated by their volume
readings, open the clamps on both containers and equilibrate the
bags to ensure that each product meets the minimum volume
requirements. Close the clamps to the storage containers when
equilibration is complete.
Note: Do not pull unnecessarily on the containers as they are
hanging from the scales.
ANUAL
with Hang Product Prompt
07-19-01-518 – January 2009