Operating and
Storage
Requirements
7-4
S
O
YSTEM
PERATOR
PECIFICATIONS
Table 7-2 provides the operating and storage requirements for
the A
LYX
Temperature Ranges
Operating*
Shipping and Storage
Altitude Maximum
Operating Conditions
Humidity Ranges
Operating*
Shipping and Storage
*Mechanical and safety system operations tested and confirmed at these
temperature and humidity extremes.
**With installation of high altitude compressor assembly kit.
Table 7-2 – A
CAUTION: Do not turn the instrument ON if condensation is
visible on the exterior of the instrument.
CAUTION: Do not operate the instrument outside of its
specified operating range. If the instrument has been stored
beyond its specified operating range, allow it to remain running
with the STOP test screen displayed for at least one hour to
stabilize to conditions before use.
CAUTION: Ensure that no devices other than the pressure cuff
and bar code scanner (if used) are attached to the connection
ports on the back of the A
connected to the A
'
M
S
ANUAL
instrument.
Fahrenheit
60 to 90° F
-40 to 158º F for up to
15 weeks
Feet
6,500
** up to 7,500
10 to 90% RH – Non-condensing
10 to 95% RH – Non-condensing
Instrument Operating/Storage
LYX
Requirements
instrument while a donor is
LYX
Component Collection System.
LYX
Centigrade
15 to 32° C
-40 to 70° C for
up to 15 weeks
Meters
2000
** up to 2286
07-19-01-518 – January 2009