07-19-01-518 – January 2009
A
C
LYX
OMPONENT
C
HAPTER
31. Remove the paper strip from the donor line and set aside.
Note: Do not place the needle and donor line on the
anticoagulant scale or on top of the anticoagulant container.
Placing anything other than the anticoagulant container on the
anticoagulant scale may cause alarms.
32. Close the clamps on the donor line and the line leading to
the sampling pack.
33. Ensure the kit has been properly loaded.
Kit Properly Loaded
WARNING: If fluid or particulate matter is present inside
the tubing lines, cassette or storage containers do not use
the disposable kit.
WARNING: Do not remove the needle from its protective
cover until it is time to perform venipuncture.