Press the cassette gently into place. Ensure the cassette fits
into all cassette guides on the instrument.
CAUTION: Press on the rigid outer ridges of the cassette only.
Excessive pressure on the inner portion of the cassette may damage
the cassette.
Cassette Properly Installed in Cassette
25.
Place the donor and anticoagulant lines securely in the
channels directly in front of the donor safety clamp. The air
detectors are located within the left channel.
Donor and ACD Lines Placed in Air
WARNING: The air detector housing and tubing channel
must be dry for proper function.
26.
Check all tubings to ensure they are clear of the cassette door
latches.
C
S
OLLECTION
YSTEM
4A – A
2RBC-LR C
LYX
Compartment
Detectors
O
'
M
PERATOR
S
ANUAL
P
OLLECTION
ROCEDURE