• A reduction of flow through the leukoreduction filter occurs
during the automated leukoreduction filtration process.
• The leukoreduction filter fixture was open while the product
was filtering.
In some instances, a product exemption message will be
displayed on the procedure information screen that instructs the
operator to check for product quality. Whenever one of these
messages appears on the procedure information screen, the data
obtained from these procedures should be excluded from
validation and quality control data. The product exemption
messages are defined as follows:
check absolute red cell content/volume for product acceptability
per institutional guidelines.
check absolute red cell content/volume and check for the
presence of residual leukocytes per institutional guidelines.
check plasma volume for product acceptability per institutional
guidelines.
check absolute red cell content of red cell product and to check
all product volumes for product acceptability per institutional
guidelines.
check all product volumes and the red cell product for absolute
red cell content, and to check the red cell product for the
presence of residual leukocytes per institutional guidelines.
ANUAL
Testing should be performed to
Testing should be performed to
Testing should be performed to
Testing should be performed to
Testing should be performed to
07-19-01-518 – January 2009