A
C
C
LYX
OMPONENT
OLLECTION
C
4A – A
2RBC-LR C
HAPTER
LYX
Check Button
Prime Icon
4A-30
S
O
'
M
YSTEM
PERATOR
S
P
OLLECTION
ROCEDURE
7.
Touch the check button to verify all cannulae have been
broken. The animated prime icon will be displayed on the
touch screen.
WARNING: Remove and discard the kit if any fluid leaks are
observed after it is primed. If you detect a leak in the kit at
any time, stop the procedure, do not reinfuse, and discard the
blood product.
Note: Donor information may be entered any time after the
operator ID and kit information is entered. This may occur either
before or after the disposable kit has been installed on the
instrument. Donor information may also be entered during the
prime sequence as described in the following section.
ANUAL
07-19-01-518 – January 2009