Fenwal 4R5707 Operator's Manual page 134

Alyx component collection system

Advertisement

A
C
C
LYX
OMPONENT
C
3 – O
HAPTER
PERATING CONSIDERATIONS
Understanding
the STOP
Button
3-68
S
O
OLLECTION
YSTEM
The STOP button is located under the touch screen and can be
used to end a procedure. Refer to the figure below.
The STOP button is best used when there is a need to stop the
centrifuge. It is not advised to use the STOP button frequently
during a procedure since stopping and starting the centrifuge
creates unnecessary strain on the umbilicus. Pushing the STOP
button while the donor is connected causes the instrument to:
• Close all valves
• Stop the centrifuge
• Stop all pumps
• Close the donor safety clamp to isolate the donor from
the instrument
• Deflate the pressure cuff
'
M
PERATOR
S
ANUAL
STOP Button
STOP
Button
07-19-01-518 – January 2009