Typical Pause Screen
Start Button
Touch the green start button on the Pause Screen to return to the
Main Collection Screen.
Pause Reminder
When the pause mode is initiated, and no action is taken by the
operator, an audible alert will sound after two minutes.
Touching the mute button will silence the alarm for two minutes.
Mute Button
Typical Pause Screen with Mute Button
In certain specific instances, such as after an alarm is cleared, or
the available saline on the Infuse Saline screen reaches zero, the
audible alert will sound within 10 seconds if no action is taken.
Touch the green mute button to silence the alarm.