Seal Kit Prompt
07-19-01-518 – January 2009
A
C
LYX
OMPONENT
C
HAPTER
Sealing the Kit
After the operator verifies that the red cell product is burped, the
procedure information screen will display the seal kit prompt.
Typical Procedure Information Screen with Seal Kit Prompt
(Jumbo Product Information Screen Displayed)
To seal the kit, perform the following steps:
Red Cell Product
1.
Seal the segment tubing between the one-way check valve
and red cell storage container, immediately next to the check
valve.
2.
Seal the other tubing, between the "Y" junction and the red
cell storage container, immediately next to the "Y".*
3.
Separate the red cell storage container from the rest of the
kit. Remove and discard the clamp.
4.
Fold the segment tubing and place in the container side
slit(s).
C
S
OLLECTION
4B – A
RBC/P
LYX
LASMA
O
'
M
YSTEM
PERATOR
S
C
P
OLLECTION
ROCEDURE
ANUAL
4B-43