Download Print this page

Fenwal 4R5707 Operator's Manual page 416

Alyx component collection system

Advertisement

A
C
C
LYX
OMPONENT
OLLECTION
C
7 – S
S
HAPTER
YSTEM
IEC 60601-1-2:2001:
Table 202 – Guidance and Manufacturer's Declaration – Electromagnetic Immunity
The A
Component Collection System is intended for use in the electromagnetic environment specified
LYX
below. The user of the A
environment.
Immunity Test
Electrostatic Discharge
(ESD)
IEC 61000-4-2
Electrical fast transient/
burst
IEC 61000-4-4
Surge
IEC 6100-4-5
Voltage dips, short
interruptions, and
voltage variations on
power supply input lines.
IEC 6100-4-11
Power frequency (50/60
Hz) magnetic field
IEC 61000-4-8
Note: UT is the a.c. mains voltage prior to application of the test level.
7-10
S
O
YSTEM
PERATOR
PECIFICATIONS
Component Collection System should assure that it is used in such an
LYX
IEC 60601 Test Level
± 6 kV contact
± 8 kV air
± 2 kV for power supply
lines
± 1 kV for input/output
lines
± 1 kV differential mode
± 2 kV common mode
< 5% UT (> 95% dip in
UT) for 0.5 cycles
40% UT (60% dip in UT)
for 5 cycles
70% UT (30% dip in UT)
for 25 cycles
< 5% UT (> 95% dip in
UT) for 5 seconds
3 A/m
'
M
S
ANUAL
Compliance Level
± 6 kV contact
± 8 kV air
± 2 kV for power supply
lines
N/A
± 1 kV differential mode
± 2 kV common mode
< 5% UT (> 95% dip in
UT) for 0.5 cycles
40% UT (60% dip in UT)
for 5 cycles
70% UT (30% dip in UT)
for 25 cycles
< 5% UT (> 95% dip in
UT) for 5 seconds
10 A/m
07-19-01-518 – January 2009
Electromagnetic
Environment Guidance
Floors should be wood,
concrete, or ceramic tile. If
floors are covered with
synthetic material, the
relative humidity should be
at least 30%.
Mains power quality should
be that of a typical
commercial or hospital
environment.
Mains power quality should
be that of a typical
commercial or hospital
environment.
Mains power quality should
be that of a typical
commercial or hospital
environment. If the user of
the A
Component
LYX
Collection System requires
continued operation during
power mains interruptions,
it is recommended that the
A
Component
LYX
Collection System be
powered from an
uninterruptible power
supply.
Power frequency magnetic
fields should be at levels
characteristic of a typical
location in a typical
commercial or hospital
environment.