Check Button
Kit-Check Icon
4A-26
S
O
'
M
YSTEM
PERATOR
S
P
OLLECTION
ROCEDURE
CAUTION: Failure to follow instructions prior to venipuncture
may result in dilution of whole blood samples.
3.
Touch the check button to verify the clamp on the donor line
has been closed. The instrument will display the animated
kit-check icon on the touch screen. The screen will also
provide information about the approximate time remaining
for the kit-checks.
Typical Main Data Screen Indicating Kit-
WARNING: Opening the clamp between the sampling
container "Y" and the anticoagulant line "Y" after the system
self-checks and before venipuncture may cause
non-sterile air to enter the kit. If this occurs, discard this kit
and restart the procedure with a new kit.
ANUAL
Checks in Progress
07-19-01-518 – January 2009