Fenwal 4R5707 Operator's Manual page 379

Alyx component collection system

Advertisement

Code
Icons
10413
Scale has been
Disturbed
Air in AC Line
10414 -
Scale has been
Disturbed
10417
10418 -
Scale has been
Disturbed
10427
10433
Scale has been
Disturbed
Air in AC Line
10434 -
Scale has been
Disturbed
10437
10440
Non-Recoverable
Alarm Screen
10441
Scale has been
Disturbed
-
10445
10504
Scale has been
Disturbed
IVD Limit
Exceeded
10506
Scale has been
Disturbed
-
10510
07-19-01-518 – January 2009
A
LYX COMPONENT COLLECTION
Meaning
The system has detected an empty
container.
The system has detected an empty
container.
The system has detected
fluctuations in the weigh scale
readings.
The system has detected that the
AC container has been removed
from the AC scale. As a result,
there may be air in the anticoagulant
line.
The system has detected that a
container was removed from a scale.
The system has detected an issue
with the weigh scales.
The system has detected an issue
with the scale calibration.
The system has detected that the
IVD limit have been exceeded.
The system has detected
fluctuations in the weigh scale
readings.
S
YSTEM
C
HAPTER
Resolution
- Check that the container that is highlighted on
the screen is properly installed on the scale.
- Follow troubleshooting steps outlined for
alarm 7025.
- Resume the procedure.
- Check that the container that is highlighted on
the screen is properly installed on the scale.
- Resume the procedure.
- Check that the container that is highlighted on
the screen is properly installed on the scale.
- Resume the procedure.
- Ensure the AC container is placed correctly on
the AC scale.
- Ensure the line from the anticoagulant
container to the cassette has no air.
- Ensure the anticoagulant line from the cassette
to the "Y" junction on the donor line has no air.
Note: Small air bubbles in the AC line
dislodged from cassette surfaces will be drawn
into the disposable during normal operation of
the instrument.
- Check that the container that is highlighted on
the screen is properly installed on the scale.
- Resume the procedure.
- Turn the instrument OFF, then back ON to
start a new procedure.
- If problem persists, contact the Hotline.
- Check that the proper weights are being used
to calibrate the scales.
- Refer to Performing Daily Set-up section in
Chapter 3 for additional information.
- If problem persists, contact the Hotline.
- Check that all containers are hanging properly
on the scales.
- Resume the procedure.
- If the IVD limit Exceeded Alarm is the
result of multiple attempts to clear air from
the donor line in response to alarm code
7025, terminate the procedure and DO NOT
return fluids to donor.
- Check that the container that is highlighted on
the screen is properly installed on the scale.
- Resume the procedure.
O
'
M
PERATOR
S
ANUAL
5 – T
ROUBLESHOOTING
5-29