07-19-01-518 – January 2009
A
C
LYX
OMPONENT
WARNING: Use only clear solutions.
WARNING: Sterile docking of any kit components,
including needles/fistulas, to the A
not been validated and is not recommended.
Exception Note: A sterile connecting device may be
used in association with Single Unit Recovery, when post-
procedure product transfer from the Red Cell Container
to a final Red Cell Storage Container must be performed.
CAUTION: Allow the kit (and separate solutions, if applicable)
to reach room temperature before use.
CAUTION: If the disposable kit is not primed, it should be
used within 8 hours after being loaded on the instrument.
However, once the cannulae have been broken on a closed kit, or
once kit-checks have started on a functionally closed kit, the
procedure must begin within 1 hour. After this time, the
instrument will alarm and prompt the operator to remove the kit.
The operator must press the STOP button, end the procedure
and remove the kit. A new procedure may then be started.
CAUTION: A disposable kit may only be reloaded as long as
the cannulae have not been broken, saline has not been spiked, if
applicable, and the sterility barrier(s) are intact. Otherwise if the
one-hour limit is exceeded, the kit must be discarded.
CAUTION: Press on the rigid outer edges of the cassette only.
Excessive pressure on the inner portion of the cassette may
damage the cassette.
CAUTION: Do not strip the segment tubing below the
leukoreduction filter until separated from the filter.
Procedure-Related Warnings and Cautions
WARNING: Complete instructions for use should be
followed for every collection procedure. The A
Component Collection System should be operated only by a
qualified individual who has been trained in the use of the
instrument.
WARNING: Do not remove the needle from its protective
cover until it is time to perform venipuncture.
WARNING: Do not leave the instrument unattended
during a collection procedure.
WARNING: Handle and dispose of all biologically
contaminated materials according to local and national
government requirements.
C
S
O
OLLECTION
YSTEM
C
HAPTER
disposable kit has
LYX
LYX
'
M
PERATOR
S
ANUAL
1 – I
NTRODUCTION
1-15