Fenwal 4R5707 Operator's Manual page 414

Alyx component collection system

Advertisement

A
C
C
LYX
OMPONENT
OLLECTION
C
7 – S
S
HAPTER
YSTEM
7-8
S
O
YSTEM
PERATOR
PECIFICATIONS
To reduce the risk of EMI, follow these recommendations:
• Do not turn on or use hand-held personal communications
devices, such as mobile two-way radios or cellular phones,
near the instrument. If these devices need to be used, follow
the "recommended separation distance" as shown in the
following tables.
• In the case of unexplained EMI, consider the locations of
nearby transmitters, such as radio or TV stations. You may
have to move the instrument or place shield material between
the transmitter and the instrument.
• Be aware that modifying the instrument or adding accessories
or components not specifically authorized by Fenwal may
make the instrument more susceptible to interference from
radio waves.
• The following cables, accessories, and transducers have been
approved for use with the A
System and comply with current EMC standards:
1.
2.
WARNING: When servicing the instrument, use only
replacement components, cables, and accessories
authorized by Fenwal and be sure to replace all shields,
covers, screws, and gaskets in their exact locations as
described in the service manual. Failure to do so may result
in increased emissions or decreased immunity of the
instrument.
WARNING: The A
should not be used adjacent to, or stacked with, other
medical electrical equipment. However, if it is necessary to
use the instrument close to other equipment, the instrument
should be observed to verify normal operation in the
configuration in which it is intended to be used.
'
M
S
ANUAL
RS232 cable
Barcode scanner
Component Collection System
LYX
Component Collection
LYX
07-19-01-518 – January 2009

Advertisement

loading

This manual is also suitable for:

4r5725