Removing the Kit
Unload Kit
Prompt
Check Button
Procedure
View
Button
New Procedure
Button
4B-50
S
O
YSTEM
PERATOR
C
P
LASMA
OLLECTION
After touching the check button to confirm the products have been
weighed, the unload kit prompt is displayed on the Procedure
Information Screen.
Typical Procedure Information Screen with Unload Kit
To remove the kit, perform the following steps:
1.
Open the cassette compartment door and centrifuge door.
2.
Remove the separation chamber completely from the
centrifuge compartment.
3.
Remove the cassette from the cassette compartment.
4.
Remove the lines from the air detectors and close the
cassette compartment door.
Remove the rest of kit from the instrument. Properly dispose of
the kit according to standard operating procedures. Touch the
check button to verify the kit has been removed from the
instrument.
When the operator confirms that the kit is unloaded, the
instrument checks that the kit has been removed. If not, the user
is prompted to remove the kit again. If the operator has
confirmed the kit has been removed, the operator is given access
to the procedure view button and new procedure button.
Note: Record the procedure data before touching the new
procedure button. The procedure view button provides summary tabs
for additional information regarding the donor, procedure results,
kit and accessory items that may have been used, and any alarms
'
M
S
ANUAL
ROCEDURE
Prompt (Split Plasma Product)
07-19-01-518 – January 2009