Alyx component collection system

Advertisement

A
C
C
LYX
OMPONENT
OLLECTION
C
4C – A
2RBC-NLR C
HAPTER
LYX
Check Button
4C-8
S
O
'
M
YSTEM
PERATOR
S
P
OLLECTION
ROCEDURE
Note: The kit code must be scanned or manually entered before
the kit installation check. The operator may change the kit code
number at any time before kit installation. If the code is manually
re-entered after the kit information has been scanned, any
previous lot, expiration date, etc., will be cleared for that kit.
Note: If the kit code is entered manually, the instrument does
not automatically record the lot number and expiration date. The
operator must ensure that the kit has an acceptable expiration
date.
7.
Touch the highlighted kit code field on the overlay and the
alphanumeric keypad for kit information will be displayed.
Typical Alphanumeric Keypad for Kit Information
Enter the appropriate kit code and touch the check button.
8.
Note: Touch the kit lot number and/or expiration date(s) entry
fields if these parameters are desired. Individual keypads will be
displayed after the parameter is touched. Enter the information
and touch the check button after each entry is made.