A
C
C
LYX
OMPONENT
OLLECTION
C
4A – A
2RBC-LR C
HAPTER
LYX
Disconnecting
the Donor
Disconnect
Donor Prompt
Processing Icon
4A-42
S
O
'
M
YSTEM
PERATOR
S
P
OLLECTION
ROCEDURE
After the final return phase is complete, the instrument provides an
audible tone and advances to the Procedure Information Screen
where the disconnect donor prompt is displayed. The instrument will
also automatically proceed to process the red cell product, as
indicated by the animated processing icon displayed on the touch
screen.
Typical Procedure Information Screen
During Product Processing
To disconnect the donor, follow these steps:
1.
Close the clamp on the donor line.
2.
Remove the pressure cuff from the donor's arm.
3.
Remove the needle while simultaneously sliding it into the
needle protector. Ensure the needle is completely covered by
the needle protector.
4.
Apply pressure to the venipuncture site to stop blood flow.
5.
Follow institutional guidelines for venipuncture site care and
release the donor according to applicable standard operating
procedures.
ANUAL
07-19-01-518 – January 2009