Fenwal 4R5707 Operator's Manual page 144

Alyx component collection system

Advertisement

A
C
C
LYX
OMPONENT
C
3 – O
HAPTER
PERATING CONSIDERATIONS
Automatic
Cassette
Refresh
Kit Timer
Break Cannulae
Prompt
3-78
S
O
OLLECTION
YSTEM
Closed Kit
The Automatic Cassette Refresh feature allows the operator to
load a closed system kit and perform kit installation checks up
to 8 hours before starting the procedure. Once the kit priming
sequence has been initiated, the Kit Timer will indicate the
allowable time remaining to start the procedure.
Note: The system will automatically refresh the pressure seal
between the cassette and the pump block gasket every
15 minutes. The timer is not displayed during the short 20 to
30 seconds the cassette is being refreshed or during prime.
When the kit installation checks are complete, the touch
screen will display the break cannulae prompt and will flash the
60:00 (minutes and seconds) timer.
Once the check button has been touched to confirm the cannulae
have been broken, the kit timer will begin.
Once prime has been completed, the timer will indicate the
time remaining to enter donor information, perform
venipuncture, and begin the procedure using the disposable kit
that is currently loaded, installed and primed.
'
M
PERATOR
S
ANUAL
Typical Screen with Break Cannulae
Prompt and Flashing Timer
07-19-01-518 – January 2009