Fenwal 4R5707 Operator's Manual page 380

Alyx component collection system

Advertisement

A
LYX COMPONENT COLLECTION
C
5 – T
HAPTER
ROUBLESHOOTING
Code
Icons
10511
Check
Venipuncture
Open Clamp
Check for Kink in
Line
10512 -
Check for Kink in
Line
10514
1051 -
Scale has been
Disturbed
Check Kit for
10518
Leaks
10521 -
Scale has been
Disturbed
Check for Kink in
10524
Line
11001 -
Non-Recoverable
Alarm Screen
11015
11017 -
Non-Recoverable
Alarm Screen
11021
11022
Non-Recoverable
Alarm Screen
11023
Check Centrifuge
11024
Non-Recoverable
Alarm Screen
11025
Non-Recoverable
Alarm Screen
11027
Non-Recoverable
Alarm Screen
5-30
S
O
'
YSTEM
PERATOR
S
Meaning
The system has detected a flow rate
that was not expected.
The system has detected a flow rate
that was not expected.
The system has detected a potential
leak in one of the containers.
The system has detected a flow rate
that was not expected.
The system has detected a software
error.
The system has detected an issue
with the power supply voltage.
The system has detected an issue
with the leak detector.
The system has detected a leak in
the centrifuge compartment.
The system has detected an issue
with the power supply voltage.
The system has detected a software
error.
The system has detected excessive
pressure in the cuff.
M
ANUAL
- Check donor line for kinks or occlusions.
- Check needle placement.
- Check donor's arm.
- Encourage donor to squeeze.
- Decrease draw/return rate.
- Check the donor line; ensure that the clamp is
open.
- Check for kinks in the tubing to and from the
filter.
- Check for kinks in the tubing from the RBC
container to the cassette.
- Check for kinks in the tubing from the
preservation solution container to the cassette.
- Resume the procedure.
- Check for kinks in the tubing that is highlighted
on the screen.
- Resume the procedure.
- Check for leaks in the container that is
highlighted on the screen.
- If no leak is found, resume the procedure.
- Terminate the procedure without
reinfusion if a leak is found.
- Check that the container that is highlighted on
the screen is properly installed on the scale.
- Resume the procedure.
- Turn the instrument OFF, then back ON to
start a new procedure.
- If problem persists, contact the Hotline.
- Turn the instrument OFF, then back ON to
start a new procedure.
- If problem persists, contact the Hotline.
- Turn the instrument OFF, then back ON to
start a new procedure.
- If problem persists, contact the Hotline.
- Turn the instrument OFF, then back ON to
start a new procedure.
- If problem persists, contact the Hotline
- Turn the instrument OFF, then back ON to
start a new procedure.
- If problem persists, contact the Hotline.
- Turn the instrument OFF, then back ON to
start a new procedure.
- If problem persists, contact the Hotline.
- Turn the instrument OFF, then back ON to
start a new procedure.
- If problem persists, contact the Hotline.
07-19-01-518 – January 2009