Fenwal 4R5707 Operator's Manual page 208

Alyx component collection system

Advertisement

A
C
C
LYX
OMPONENT
OLLECTION
C
4A – A
2RBC-LR C
HAPTER
LYX
Monitoring
Collection
Donor Draw
Icon
Donor Return
Icon
4A-38
S
O
'
M
YSTEM
PERATOR
S
P
OLLECTION
ROCEDURE
During the procedure, blood from the donor is withdrawn and
returned in several short cycles; each cycle includes a draw phase
and a return phase. The icon in the task zone indicates whether the
instrument is in a draw or return phase.
The first return cycle will return all contents from the plasma
container to remove saline from the system. Packed red cells will
be collected into the red cell container throughout the procedure.
Plasma is temporarily collected in the plasma container during the
draw phase and is returned with some saline at each subsequent
return cycle. When the last draw phase begins, enough whole
blood is drawn to reach the red cell collection target. Once the
target volume has been reached, any remaining blood products are
returned.
WARNING: Do not leave the instrument unattended during
a collection procedure.
WARNING: Remove and discard the kit if any fluid leaks are
observed after it is primed. If you detect a leak in the kit at
any time, stop the procedure, do not reinfuse, and discard the
blood product.
WARNING: If the instrument is accidentally turned OFF
during a procedure, immediately isolate the donor from the
instrument by closing the clamp on the donor line. Do not
turn the instrument ON with the donor connected.
CAUTION: The system monitors the weight of containers
hanging on the scales. Do not lift or pull on the containers, as this
may cause alarms.