Fenwal 4R5707 Operator's Manual page 82

Alyx component collection system

Advertisement

A
C
C
LYX
OMPONENT
C
3 – O
HAPTER
PERATING CONSIDERATIONS
Scale Pass Icon
3-16
S
O
OLLECTION
YSTEM
3.
Remove the 1000 gram weight when the button animation
stops, and the scale pass icon replaces the 1000 gram
calibration button. The 0 gram calibration button will be
highlighted.
4.
Touch the highlighted 0 gram calibration button. The button
animation will indicate the instrument is calibrating the
scale to 0 grams.
5.
Follow steps 1 thru 4 as outlined above to calibrate any
other scales indicated on the scale calibration screen.
Continue calibrating scales until all scales on the calibration
screen have been successfully calibrated.
'
M
PERATOR
S
ANUAL
Typical Scale Calibration with 0 grams
Typical Scale Calibrating to 0 grams
07-19-01-518 – January 2009