Check Button
Burp Product
Prompt
4D-26
S
O
YSTEM
PERATOR
U
R
INGLE
NIT
ECOVERY
The operator should touch the check button after the donor has
been appropriately disconnected and the processing phase is
complete.
Typical Procedure Information Screen – Confirm Remove
Burping the Product
The burp product prompt will be displayed on the touch screen.
Typical Procedure Information Screen with Burp Product
To burp the red cell product, follow these steps:
1.
Confirm that the clamp to the empty storage container is
closed.
2.
Remove the paper strip from the filled red cell storage
container segment tubing, if desired.
'
M
S
ANUAL
P
ROCEDURE
Needle (SUR)
Prompt (SUR)
07-19-01-518 – January 2009