Download Print this page

Fenwal 4R5707 Operator's Manual page 355

Alyx component collection system

Advertisement

A
C
C
S
O
'
M
LYX
OMPONENT
OLLECTION
YSTEM
PERATOR
S
ANUAL
C
5 – T
HAPTER
ROUBLESHOOTING
The operator is prompted to disconnect the donor, turn the
instrument OFF, and then remove the kit. The operator cannot
resume the procedure, automatically return fluids, or process any
collected red cell product. The operator is unable to remove the
kit without first turning the instrument OFF.
Note: Ensure the instrument remains turned OFF for at least
5 seconds before attempting to turn it back ON.
WARNING: If the instrument is accidentally turned OFF
during a procedure, immediately isolate the donor from the
instrument by closing the clamp on the donor line. Do not
turn the instrument ON with the donor connected.
WARNING: Do not open the cassette compartment door
while a donor is attached to the instrument.
There is one non-recoverable alarm that requires the operator to
turn the instrument OFF, but displays a different type of alarm
screen. An example is provided below.
Typical Non-Recoverable Alarm Screen
Requiring Power Cycling
07-19-01-518 – January 2009
5-5

Advertisement

loading

This manual is also suitable for:

4r5725