Fenwal 4R5707 Operator's Manual page 292

Alyx component collection system

Advertisement

A
C
C
LYX
OMPONENT
OLLECTION
C
4C – A
2RBC-NLR C
HAPTER
LYX
Saline Icon
4C-12
S
O
'
M
YSTEM
PERATOR
S
P
OLLECTION
ROCEDURE
7.
Hang the saline container on the saline scale. The icons on
the saline container and the saline scale match.
8.
Leave the paper strip on the red cell storage container
segment tubing, if desired and hang the containers on the
storage container hanger located on the front of the
instrument. Ensure all clamps to the red cell storage
containers are open.
Note: If the instrument is positioned on a cart in such a manner
that the red cell storage containers cannot hang freely from the
hanger, they may be hung from a hook attached to the front or
side of the cart. Hang both containers from the same hook.
ANUAL
Saline Container on the
Saline Scale
07-19-01-518 – January 2009