CAUTION: The touch screen is the main input device for the
instrument. Do not touch the screen during operation except to
interact with the instrument.
CAUTION: Do not operate the instrument if the display is
blank, unreadable, or if the instrument does not respond to the
touch screen commands. If the display becomes blank during a
procedure, press the STOP button to isolate the donor from the
instrument.
CAUTION: Do not operate the instrument unless both fan
filters are properly installed.
CAUTION: Do not drape blankets or other materials over the
instrument during operation. This may obstruct ventilation and
cause the instrument to overheat.
CAUTION: The battery will not resume the procedure in the
case of power loss due to accidental switching of the power
switch to OFF during a procedure.
CAUTION: Do not use solvents, abrasive cleaners, or a greater
than 10 percent bleach solution to clean any part of the
instrument. Use only recommended cleaning solutions. Refer to
the cleaning instructions for each area to determine which
solutions and cleaning cloths are recommended.
CAUTION: Do not use bleach to clean or disinfect the air
detectors. Using bleach on these parts may cause damage.
CAUTION: Do not spray fluids directly onto the back of the
instrument housing or near the saline scale. If fluid enters the
inside of the instrument, contact your qualified service
representative. Do not attempt to clean a spill in those locations.
CAUTION: Only operate the instrument with a Fenwal
approved power cord within the United States.
CAUTION: Outside the United States, only operate the
instrument with a Fenwal approved cord or a power cord that
meets the specifications outlined in this manual.
CAUTION: Ensure the instrument is level prior to starting a
collection procedure. Ensure the instrument is level each time it
is repositioned or relocated.
CAUTION: Do not turn the instrument ON if condensation is
visible on the exterior of the instrument.
CAUTION: The instrument should be turned OFF at least
once daily, or remain OFF if not in use. This allows the
instrument to perform automated self-checks needed to evaluate
system integrity when it is turned back ON.