Download Print this page

Fenwal 4R5707 Operator's Manual page 389

Alyx component collection system

Advertisement

Pause Button
07-19-01-518 – January 2009
A
LYX COMPONENT COLLECTION
7.
Install a new kit and proceed to kit checks and prime. If the
noise returns, turn the instrument OFF, remove the kit, and
contact the Hotline.
Fluid in Cassette and/or Tubings Before Use
WARNING: If fluid or particulate matter is present inside
the tubing, cassettes, or storage containers of the disposable
kit prior to kit installation, discard the kit.
Whenever fluid is found in the tubing, cassettes, or storage
containers before the disposable kit has been used, the kit should
be discarded. Obtain a new disposable kit before beginning a
new procedure.
Unsuccessful Venipuncture
Use the following instructions if the initial venipuncture is
unsuccessful.
1.
Touch the pause button to stop the procedure if it has been
started.
2.
Close the clamp on the donor line.
3.
Disconnect the donor. Follow appropriate institutional
guidelines for venipuncture site care.
WARNING: Sterile docking of any kit components,
including needles/fistulas, to the A
not been validated and is not recommended.
Exception Note: A sterile connecting device may be
used in association with Single Unit Recovery, when post-
procedure product transfer from the Red Cell Container
to a final Red Cell Storage Container must be performed.
Fluid Spills or Leaks
Discontinue the procedure without fluid return immediately and
disconnect the donor. Refer to the instructions in Chapter 6 to
clean blood or fluid spills.
WARNING: Do not attempt to reinfuse blood components
following fluid leaks or spills, when the instrument is
operating with unexpected noise or vibration, or when a
procedure is ended due to persistent redness in the plasma
tubing line coming from the centrifuge. In these instances,
the products must be discarded.
S
O
YSTEM
PERATOR
C
5 – T
HAPTER
ROUBLESHOOTING
disposiable kit has
LYX
'
M
S
ANUAL
5-39

Advertisement

loading

This manual is also suitable for:

4r5725