Dmc Pressure And Flow Board; Figure 7-20. Flow Pressure Board, Disconnect Dmc Cable - Thermo Scientific 48iQ Instruction Manual

Carbon monoxide analyzer

Advertisement

DMC Pressure
and Flow Board
Thermo Scientific
Use the following to replace the DMC pressure and flow board.
Equipment required:
Hex drive, 7/16
1. Turn instrument OFF, unplug power cord, and remove the cover
(Figure
2–1).
2. Unplug cables from the pressure and flow board.
Figure 7–20. Flow Pressure Board, Disconnect DMC Cable
3. Disconnect plumbing.
4. Using 7/16 hex drive, unfasten four #6-32 socket cap head screws.
Servicing

DMC Pressure and Flow Board

48iQ Instruction Manual
7-17