9 Installation
Minimum operating space requirements
Depth
Width
Height
9.2.3 Fixing hardware and tools
The following tools and hardware are required to install and set up the device. These are
not included in the delivery, unless otherwise stated.
Fixing hardware
NOTICE! Mounting bolts for floor and wall fastening are not included in the
delivery. The fixing hardware used to fix the device to the wall must be of correct
type for the wall.
The type and length of hardware to be used depend on the wall material and floor
material where the device is to be fixed. The fixing hardware must be selected so that
they withstand at least 1500 N pull-out force.
WARNING! Ensure that the mounting bolts, the wall and the floor the device is
attached to, can withstand at least 1500 N pull-out force.
Installation tools
• Electric drill
• Spanners size 13 & 17 mm
• Allen keys 3, 5 & 6 mm
• Flat head screwdriver
• Spirit level
• Pliers and wire cutter
• Scissors/knife
• Torque wrench
56
Minimum:
1385 mm / 54.5 in
Recommended
minimum:
1630 mm / 64.2 in
Minimum:
1375 mm / 54 in
Recommended
minimum:
1610 mm / 63.4 in
Minimum:
2080 mm / 81.9 in
Recommended
minimum:
2467 mm / 97.1 in
min. 715 mm / 28.1 in
min. 660 mm / 26 in
rec. 950 mm / 37.4 in
ORTHOPANTOMOGRAPH
™