6 Panoramic and 3D calibrations and adjustments
84
NOTICE! Lasers can be turned on by pushing the light
button. Lasers stay on approx. 30 seconds.
6. Briefly press the exposure button (no X-rays are generat-
ed) to acknowledge the adjustment has been carried out.
The rotation and Y-position are also saved.
6.5.11
3D pixel calibration
1. Remove all phantoms (if attached).
2. Select the 3D Pix program.
3. Press Patient Positioning.
4. Take an exposure.
5. The resulting image informs when the calibration is
passed.
6.5.12
Panoramic pixel calibration
NOTICE! The pixel calibration results are sensor specific.
If the x-ray unit is equipped with separate panoramic and
cephalometric sensors, the cephalometric sensor cannot
be used for panoramic imaging without re-calibration (and
vice versa).