Technician's instructions SURGmatic S201 L - 1.009.0470
2 Disassembly | 2.1 Disassembly of the SURGmatic S201 L

2 Disassembly

2.1 Disassembly of the SURGmatic S201 L

2.1.1 Removing the head from the elbow sleeve

▶ Pull the spray tube ② off at the taper sleeve ③ and remove it from the head
①.
▶ Open the clamping ring ④ by turning clockwise and remove the head ① from
the elbow sleeve ⑤.

2.1.2 Disassembling the head casing

Tool
Name
Screwdriver (Mat. no. 0.308.0004)
Screwdriver bit (Mat. no. 0.410.0784)
Circlip pliers (Mat. no. 0.411.2421)
Wrench (Mat. no. 0.411.2371)
Universal bit holder
(Mat. no. 1.002.4577)
Torque handle (Mat. no. 1.002.4578)
100 Ncm