9.8 Complete device calibration

After the installation of the device and during the annual maintenance, a complete
calibration procedure must be performed. See chapter
information on the unit calibration.
Perform the calibrations in the order as described in this chapter.
After the calibrations have been completed, the QC images need to be taken and stored
for the user as reference.

9.8.1 Accessing the calibrations

A complete list of calibration programs, required to be performed after the unit
installation, can be found in the PIN code protected Service menu.
NOTICE! The calibrations programs in the Service menu are intended for service
technicians only, and not to be performed by the end user of the unit.
NOTICE! The device's country selection needs to be configured correctly before
performing the calibrations. See chapter
1. Turn the device on and establish a connection to the workstation. See chapters
Powering on the device
2. Select Settings menu on the GUI.
3. Select Service menu and enter PIN code 0612.
4. The GUI informs that additional options have been added under the Service menu.
5. Select Calibrations under the Service menu to see the list of all calibration
programs.
ORTHOPANTOMOGRAPH
and
Preparing the device for imaging
OP 3D
Preparation for Calibration
Device
configuration.
for more information.
9 Installation
for
73