Operating Precautions; Cleaning The Sample Windows - Thermo Scientific Nicolet iS5 LS User Manual

Table of Contents

Advertisement

2
Operating the Instrument

Operating Precautions

10
Nicolet iS5 LS User Guide
If a component status is red (failed), see the
instrument user guide or contact us for assistance.
Related Topics
Troubleshooting the Instrument
Operating the Instrument
WARNING Prevent fire and explosion. The infrared source inside the instrument is an
ignition source. If you are using volatile solvents, provide a fume hood or other active
venting system that is free of spark and other ignition sources and prevents flammable
vapors from collecting in the atmosphere surrounding the instrument.
To prolong the life of your instrument and eliminate the possibility of spectral interference
caused by volatile solvent vapors, follow these precautions:
• When cleaning sample windows, avoid using excessive force or you may damage or crack
the windows.
• Do not block the vents at the back of the analyzer. Electronic components can overheat if
these vents are blocked.
• Do not place containers of liquids on top of the analyzer. Spilled liquids can enter the
vents or other openings and cause damage.
• Do not use chlorinated solvents or materials that may produce HCl or HF vapors with or
near the analyzer. These solvents may severely damage the instrument.
• Do not leave solvent on the sample windows for longer than 2 minutes.
NOTICE Read the safety guide that came with this instrument before operating the
instrument.
"Troubleshooting the
Instrument" section of the
Thermo Scientific