GE Revolution CT User Manual page 668

Tomography scanner
Table of Contents

Advertisement

Table 2: Acquire Scan Operator Messages
Number English Operator Message
Data Collection Hardware Unavailable.
Please perform a Scan Hardware Reset from the Service
1
Tools menu.
If problem persists, contact your service representative.
Data Collection Hardware configured incorrectly. Slip ring
2
miswired.
Please contact your service representative.
Data Collection Hardware Unavailable.
Please select Done Scanning and perform a Scan Hard‐
3
ware Reset from the Service Tools menu.
If problem persists, contact your service representative.
The scan database is corrupted.
If you continue to use the system you might propagate the
4
corruption. Please perform a System Restart from the
Mode menu in order to automatically recover the database.
If problem persists, contact your service representative.
Unable to reserve scan data.
5
Please perform a System Restart from the Mode menu.
Acquired data is corrupt.
6
Scanning will be disabled.
Please contact your service representative.
Failed to start exposure due to system problem. Scan time
out.
7
If problem persists, please contact your service representa‐
tive.
Actual dose information for the scanned group unavailable,
recorded dose has been estimated from the scan prescrip‐
8
tion.
If problem persists, contact your service representative.
Firmware reset occurred. Please wait until reset is complet‐
9
ed.
10
Landmark not set.
11
Prescribed interscan delay not met.
Real Time Display error.
Displayed indication may be inconsistent with actual sys‐
12
tem state.
Please contact your service representative.
System configuration data is invalid or corrupt.
13
Please perform a System Restart from the Mode menu.
If problem persists, contact your service representative.
System could not detect ECG Triggers.
14
Please check ECG lead connection.
System detected software error during scanning.
15
Please perform a System Restart from the Mode menu.
If problem persists, contact your service representative.
System is recovering after serious error, please check sus‐
16
pended image reconstructions in the queue.
668
Revolution CT User Manual
Direction 5480385-1EN, Revision 1