Enhance is available in the
slice thickness is 0.625 mm or 1.25 mm. The interval for these slice thicknesses must be equal
to the slice thickness (0.625 or 1.25) or one half the slice thickness (0.312 or 0.625).
E " displays on the image next to the annotation for to indicate IQ Enhance is enabled for the
An "
acquisition.
NOTE:
IQ Enhance is not compatible with Axial, Cine or Cardiac scan types.

6.1.7 Quality Check (QC) Images

QC images are non-diagnostic, preview images generated by the system to check if the scan
acquisition was successful, prior to full IQ images. QC images are displayed in the two
viewports on the scan monitor. QC images have an annotation at the bottom to indicate that
they are not for diagnostic use. QC images are only generated once during a scan series, and
are not saved or available after the scan series is acquired.
QC Images will be auto-generated with the Primary Recon for the following scan types: Axial ,
Cardiac Axial, Helical and Cine.

6.1.8 Flip / Rotate

The reconstructed orientation of an image can be changed from the default mode by selecting
one of the following flip/rotate options in the
None
FLR (Flip Left Right)
FTB (Flip Top Bottom)
FLR / FTB (Flip Left Right / Flip Top Bottom)
NOTE:
Flip / Rotate options cannot be selected for images being acquired in the decubitus
patient positions.
290
Primary Recon and Secondary Recon parameter collections, if the
Illustration 23: Image Enhance images
Primary Recon and Secondary Recon collections.
Revolution CT User Manual
Direction 5480385-1EN, Revision 1
6 Recon Parameters

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Revolution CT and is the answer not in the manual?

Questions and answers

Greg Lesco
February 27, 2025

need a part number for the Integrated Cardiac Patient Monitor, patient grounding strap + cable

Yaya
March 14, 2025

How do I clear Ge Revolution collision sensor

1 comments:
Mr. Anderson
March 14, 2025

To clear the collision sensor on a GE Revolution CT, remove any obstructions detected by the sensors located beneath the front and back of the table. Once the area is clear, you can continue positioning the table.

This answer is automatically generated

Subscribe to Our Youtube Channel

Table of Contents

Save PDF