NOTE:
Flip/Rotate options are only available after service checks that the viewing stations
will display the data correctly.

6.1.9 Copy Forward

Copy Forward Patient Orientation, Patient Position and Anatomical Reference is a feature that
adjusts these three bundled parameters in a scanning protocol based on the values in the
previous series. This feature is located in Protocol Management.
NOTE:
Copy Forward is set at the destination or recipient series.
For example, if Copy Forward was turned "On" in Series 2 of a protocol, and during
scan Patient Orientation, Patient Position and Anatomical Reference were changed
from Head First, Supine, SN to Feet First, Prone, OM for Series 1, these three
settings will be automatically copied for scanning Series 2.
Similar to Copy Patient Orientation / Position, Copy Forward works within the same
patient exam.
The Copy Forward feature can also be used to duplicate (reuse) the following parameters within
series from primary reconstruction to secondary reconstructions:
Start Location
End Location
Time Between Images (Cine scan type only)
Interval (Helical scan type only)
DFOV
R/L Center
A/P Center
To activate Copy Forward, type d or D (duplicate) in the
reconstruction.

6.2 Primary Recon collection

Scan Settings window, click the Primary Recon collection.
From the
Chapter 11 Scan
Revolution CT User Manual
Direction 5480385-1EN, Revision 1
Parameter text field in the receiving
291