Scan Applications chapter; SmartPrep section for more details.
12. See the
Typical SmartPrep parameters for cardiac scans:
mA = 40–60
Monitoring Delay = 10.0 seconds
Monitoring Inter-Scan Delay (ISD) = 1.7 seconds produces an image every 2.0 seconds
Enhancement Threshold = 150
Diagnostic Delay = Auto Minimum or desired delay time
13. Arm the injector.
14. Click [Confirm Settings].
15. Give the patient the same breathing instructions as the previous series. See
more details.
16. Press Move to Scan.
17. Start the injector.
If using the Integrated Injector option: The injector is started automatically when Start
Scan is pressed.
If not using the Integrated Injector option: Simultaneously press Start the Injector and
Start Scan.
344
4
Saline for 10 seconds
5
X-ray on
Revolution CT User Manual
Direction 5480385-1EN, Revision 1
Section 1.3
2 Cardiac Workflow