Ped Head SFOV supports DFOVs up to 32 cm in diameter and uses IBO processing to
correct for beam hardening effects. It is particularly useful for infants 18 months or less
in age.
Small Head SFOV should be used for patients from 1.5 years old to 10 years old.
Ped Body SFOV supports DFOV's up to 32 cm in diameter.
Both the Ped Body and Ped Head SFOVs are limited to 42 kW. This limits the maximum
mA possible to 350 at 120 kVp.
132
Revolution CT User Manual
Direction 5480385-1EN, Revision 1
3 Pediatric and small patient scans