Sending anonymized data does not guaranty that data will be rendered anonymous in
compliance with applicable data privacy laws. Review the scan data before sharing the data
with third parties to ensure compliance with applicable privacy laws or regulations.
Raw data saved anonymously and then restored will reconstruct with patient identifying
information removed.
NOTE:
Scan must be idle when you start this procedure. Each cardiac, non-Hi Res scan file
takes approximately 5 minutes to save, if it was a 0.28 second scan. Scan times can
be longer based on the length of the exposure.
Make sure the selected scan files do not exceed the space available on the USB.
1. Connect a USB device to the USB port on the front of the media tower.
2. In File Manager, highlight any exam to activate the control panel icons.
3.
From the File Manager control panel, click the
4. Highlight the exam/series you wish to save anonymously.
5. Right click and select
Anonimized Copy or Full Copy .
6. Select
NOTE:
Make sure the USB to which you are sending the data has enough disk space to
store the data. The system does not check disk space before sending.
7. Click [Close] to exit.
8. Once the save is complete to the USB device, exit Scan Data Manager and disconnect the
USB device. Then, resume scanning.
NOTE:
See Fields removed from exams, series, or images using Full or Partial Anonymous
modes (page 19-47) to view the data that is removed.
564
Send Scan Data To .
Revolution CT User Manual
Direction 5480385-1EN, Revision 1
Scan Data Manager icon
.
8 Scan Data Manager