3.4 Operate the gantry controls

Gantry controls are described in the table, below.
172
19
AC Power Outlet – Optional GEHC supplied motorized Scanner Desktop table
Gantry and console outlets should not be used for the MedRad injector components
or ISI900. A separate grounding cable should be attached between the injector and
the CT system.
CAUTION
These receptacles are not for general use. The RPM power consumption should
not exceed 720 Watts. The Injector power consumption should not exceed 780
Watts.
DANGER
INFORMATION ON INTERNAL GANTRY COMPONENTS IS PROVIDED FOR
USER EDUCATION. THE GANTRY CONTAINS DANGEROUS VOLTAGES
AND MOVING PARTS. TO PREVENT ELECTRICAL SHOCK OR CRUSHING
INJURIES, DO NOT REMOVE COVERS OR ENTER THE GANTRY. ONLY
TRAINED, QUALIFIED SERVICE PERSONNEL MAY REMOVE GANTRY OR
OTHER EQUIPMENT COVERS.
Illustration 12: Gantry Controls
Revolution CT User Manual
Direction 5480385-1EN, Revision 1