8 Build Protocols with Notification Values by Factor
Use this procedure to set the Notification Values (NV) in user protocols for all scan groups by a
factor.
For example, if the CTDI
value for the group would be 7.5.
•
Values can by entered in increments of 0.1.
•
Values must be > 0.
•
The maximum value is 10.
1. From the image monitor, click
Mode menu, click Protocol Management , then Protocol Management .
2. From the
3. Enter your Log On Name and Password and click [OK] to continue.
Protocol Selection window, select Adult or Pediatric .
4. From the
Anatomical area, select an anatomical region.
5. From the
Protocol Selection window, select a protocol and click [Edit].
6. From the
7. Click the series task to go to the first non-scout series in the protocol.
Dose Display area, click NV CTDI
8. From the
Factor from the drop-down menu and type the desired factor for CTDI
9. Select
10. Click [Confirm Changes].
11. Click [Save] or [Save As] to return to the
Chapter 7 Dose Check
in a scan group were 5 and a factor of 1.5 was entered, the new NV
vol
.
Protocol Selection window.
Direction 5480385-1EN, Revision 1
DLP .
or
vol
Revolution CT User Manual
DLP .
and/or
vol
149