An additional trigger will display on the ECG trace.
3. To delete a trigger point on the ECG trace and to normalize the heart cycle, place the cursor
on the trigger and right-click
4. To move a trigger, place the cursor over the trigger and click and drag it to the new location.

6.8 Change heartbeat from Arrhythmia Retrigger to Reconstruct

1. Place the cursor in the ECG Editor, and right-click and select
2. The reconstruction window moves from one heartbeat to the other heartbeat.

6.9 Display secondary recon images

6.9.1 Cardiac image annotation

SSCIN (Secondary Cardiac Axial — single sector image)

6.9.2 Edited cardiac image annotation

Image where the ECG trace has been edited, are annotated with an "E" placed before the scan
type, e.g. E/SSCIN.
E/SSCIN (Secondary Cardiac Axial — single sector image)
366
Delete R-peak Trigger .
Revolution CT User Manual
Direction 5480385-1EN, Revision 1
Change Scan to Reconstruct .
6 Cardiac Recon