3.5 Run Fast Calibrations

202
NOTICE
Please refer to the Safety chapter for important safety information regarding the
use of the equipment and software on this system.
If the detectors are not at operating temperature, a message will indicate
when they are.
If there is a large shift in room temperature (±10 degrees), perform a Fast Cal
to maintain optimum IQ. Once the room temperature has stabilized, perform
another Fast Cal.
For the permissible environmental operating conditions for the CT scanner,
see Technical Reference Manual Chapter 7 Systems Specifications,
Environmental Specifications.
The system will notify you if it has been more than 24 hours since the last
Fast Cal was run.
If you experience a scan abort during Fast Cal, notify your Service Engineer.
The system will notify you if an unrecognized tube has been installed on the
system. Fast Calibration techniques are designed specifically for GE tubes
and GE cannot guarantee that the performance of the system will meet
specifications with an unrecognized tube. Fast Cal will run in
1. Clear the gantry of any objects or personnel. Any obstruction in the gantry
may lead to artifacts in scanned images.
Utilities menu, click Daily Prep , then Fast Cal .
2. From the
Service-Calibration Tools window opens.
The
Fast Cal from the drop-down and click [Run].
3. Select
4. Press [Start Scan]. Respond to all messages that display during the
calibration (approximately 15 to 20 minutes). Execute any necessary actions
indicated in the messages.
If the system detects a dirty scan window (which may be caused by a
beam obstruction, something in the gantry, contrast, dirt, or other matter
on the scan window), a message displays. Clean the scan window and
click [Retry] or click [Continue] to skip the check.
The following automated calibration check procedures occur: tube
warmup (if needed), balance check, scan window check, two channel
map scans and basis air scans. Read the screen messages for calibration
status.
5. The system returns to the
are complete. Click [Dismiss] to exit the window.
Direction 5480385-1EN, Revision 1
Service-Calibration Tools window when the checks
Revolution CT User Manual
AutoScan mode.
3 Daily Quality Assurance

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Revolution CT and is the answer not in the manual?

Questions and answers

Greg Lesco
February 27, 2025

need a part number for the Integrated Cardiac Patient Monitor, patient grounding strap + cable

Yaya
March 14, 2025

How do I clear Ge Revolution collision sensor

1 comments:
Mr. Anderson
March 14, 2025

To clear the collision sensor on a GE Revolution CT, remove any obstructions detected by the sensors located beneath the front and back of the table. Once the area is clear, you can continue positioning the table.

This answer is automatically generated

Table of Contents

Save PDF