10 Scan Using Notification Value Checking
Use this procedure to scan using Notification Value (NV) Checking.
The CTDI
or DLP value will be highlighted when the estimated dose for a scan group exceeds
vol
the NV value set for each scan group.
Adjust the scan parameters for the scan group until the exceeded NV value is below the
prescribed NV value. If the NV value is still exceeded after you click [Confirm], a confirmation
message will display. The scan cannot be confirmed until the NV value is no longer exceeded,
or you have entered a diagnostic reason for continuing the scan with the exceeded NV value.
152
Illustration 5: NV Exceeded in Dose Display
Revolution CT User Manual
Direction 5480385-1EN, Revision 1
10 Scan Using Notification Value Checking