Set The Anatomy Selection; Set The Start And End Locations; Set The Scan Field Of View - GE Revolution CT User Manual

Tomography scanner
Table of Contents

Advertisement

kV and mA Control collection, set the kV Mode to kV Assist .
2. From the
Auto Adjust will be "On" in the Primary Recon collection by default for all recons to
NOTE:
optimize the protocol's reference WW and WL based on the expected change in
image contrast and noise when kV Assist is enabled. Select "On" to use the WW
and WL values set in the Primary Recon. To change the WW and WL values, kV
mode must be set to
3. Specify the Clinical Task for the study.
4. Review the resulting kV, Noise Index or Manual mA and WW/WL values to ensure they are
acceptable.
NOTE:
Manual mA or Smart mA parameters cannot be changed when kV Assist is on. Be
sure to set these settings prior to turning on kV Assist. Secondary recons will not
automatically adjust WW and WL.

4.5.4 Set the Anatomy Selection

4.5.4.1 Set the Start and End Locations

The start and end locations are set in millimeters superior to and inferior from the anatomical
reference point. The locations designate the points of anatomy to be scanned. The easiest way
to set the start and end locations is to turn on Show Localizer and adjust the scan range
graphically. If specific locations are known, use this procedure to designate the beginning and
end points of the anatomy to be scanned.
Scan Settings window, click Anatomy Selection .
1. From the
2. For a Helical scan type, click
For Axial, Cine or Cardiac scan types, click
3. Type a beginning point.
The numbers must be preceded by the correct designation of S for superior (toward the
head) from the centering point, or I for inferior (toward the feet).
Plus + can be used for superior and minus - for inferior.
4. For a Helical scan type, click
For Axial, Cine or Cardiac scan types, click
5. Type an end point.
The numbers must be preceded by the correct designation of S for superior (toward the
head) from the centering point, or I for inferior (toward the feet).
Plus + can be used for superior and minus - for inferior.

4.5.4.2 Set the Scan Field of View

Scan Field of View (SFOV) limits how much anatomy is scanned. Use this procedure to select a
SFOV that covers the anatomy of interest. The SFOV should always be larger than the diameter
of the patient, regardless of what anatomy is being imaged.
Scan Settings window, click Anatomy Selection .
1. From the
Chapter 11 Scan
Manual kV .
Start .
Full FOV Start
End .
Full FOV End
Revolution CT User Manual
Direction 5480385-1EN, Revision 1
283

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Revolution CT and is the answer not in the manual?

Questions and answers

Greg Lesco
February 27, 2025

need a part number for the Integrated Cardiac Patient Monitor, patient grounding strap + cable

Yaya
March 14, 2025

How do I clear Ge Revolution collision sensor

1 comments:
Mr. Anderson
March 14, 2025

To clear the collision sensor on a GE Revolution CT, remove any obstructions detected by the sensors located beneath the front and back of the table. Once the area is clear, you can continue positioning the table.