CARESCAPE R860 Lung Protective Tools Quick Reference Guide Menu Adult No Alarms Insp Manual Hold Hold Breath Airway Pressure FiO2 PEEPe Peak Pressure cmH2O cmH2O cmH2O Pmax Ppeak Tidal Volume Minute Volume Respiratory Rate l/min /min 12.0 PEEP VTexp FiO2...
Page 2
Notice The materials contained in this document are intended for educational purposes only. This document does not establish specifications, operating procedures or maintenance methods for any of the products referenced. Always refer to the official written materials (labeling) provided with the product for specifications, operating procedures and maintenance requirements.
Steady State ................... .8 Using CARESCAPE R860 Clinical Decision Support Tool to Find Steady State for FRC Measurements ..9 Performing an FRC INview Procedure.
PEEP and lower pressures are intended to decrease lung injury and mortality risk . The CARESCAPE R860 offers an individualized approach to lung protective strategies with the use of FRC INview, PEEP IN view, Lung INview and SpiroDynamics.
Equipment Set-up Attach the Module Bay and Module CARESCAPE 860 INSP 1. Connect one end of the cable to the airway module 4. Connect the end of the cable to the airway module bay connection on the back of the ventilator, and bay connection, and tighten the screws.
Pmax Ppeak Tidal Volume Minute Volume Respiratory Rate l/min /min PEEP 12.0 VTexp FiO2 Tube compensation on 15:36 FiO2 Pinsp Rate Tinsp PEEP Pmax 0.90 Current Mode Standby A/C PC cmH2O /min cmH2O cmH2O CARESCAPE R860 Exhaust INSP Assembled View...
Page 7
0.90 Current Mode Standby A/C PC cmH2O /min cmH2O cmH2O CARESCAPE R860 Position of HME 4. If an HME (Heat and Moist Exchanger) is used, ensure that it is placed between the D-lite sensor and the patient ETT. Exhaust INSP...
Steady State Prior to taking an FRC measurement the patient should be in a steady state with a constant level of delivered oxygen to ensure accuracy. The patient’s gas exchange must be stable for at least 10 minutes. To ensure accuracy: 1.
Using CARESCAPE R860 Clinical Decision Support Tool to Find Steady State for FRC Measurements Note: To enter the Future/Clinical Decision Support Workspace, select the icon (as shown). If the icon is not available, then you are already in the Future/Decision Support Workspace.
Performing an FRC INview Procedure 1. Scroll to the Clinical Decision Support Workspace. 2. Select FRC. 3. Select FRC INview tab. 4. Select FRC O and set the FRC O value. Note: FRC O is adjustable to within plus or minus 10% of set FiO value.
Performing a PEEP INview Procedure 1. Scroll to Clinical Decision Support Workspace. After start is selected: 2. Select FRC. • The PEEP setting changes to the start PEEP setting 3. Select PEEP INview. • The FiO changes to the set FRC O , then returns to the set FiO to calculate FRC at the start PEEP.
The use of Lung INview and SpiroDynamics gives the clinician an enhanced view of the patient’s lung function with the addition of the intratracheal catheter. SpiroDynamics is a CARESCAPE R860 clinical decision support tool which draws a dynamic P/V curve, breath-by- breath, regardless of the volume- or pressure-controlled modes of ventilation. The catheter provides an accurate measurement of pressure delivery to the lungs by removing the resistance of the endotracheal tube from the spirometry loop.
SpiroDynamics Settings Menu 1. Scroll to Clinical Decision Support Tools. 2. Select the Spirometry view. 3. Select the SpiroDynamics tab. 4. Select the SpiroDynamics loop Field. 5. Select the settings icon in the upper right hand corner to access the settings menu. 6.
Viewing SpiroDynamics Data 1. Scroll to Clinical Decision Support Workspace. 2. Select Spirometry view. 3. Select the SpiroDynamics tab. Note: The current SpiroDynamics and Spirometry data is shown based on the SpiroDynamics settings. See SpiroDynamics settings menu for more information. 4.
Performing a Lung INview Procedure 1. Attach the intratracheal pressure sensor to the 7. Scroll to Clinical Decision Support Workspace. auxiliary pressure port on the ventilator. 8. Select FRC. 2. Access Menu. 9. Select PEEP INview tab. 3. Select System. 10.
Individualized settings can help prevent Ventilator-Induced Lung Injury and shorten time in the ICU for your patients. CARESCAPE R860 provides the ability to assess the volume and pressure in the patient’s lungs so that clinician’s can tailor the vent settings accordingly.
Page 18
GE Healthcare provides transformational medical technologies and services that are shaping a new age of patient care. Our broad expertise in medical imaging and information technologies, medical diagnostics, patient monitoring systems, drug discovery, biopharmaceutical manufacturing technologies, performance improvement and performance solutions services help our customers to deliver better care to more people around the world at a lower cost.