Table of Contents

Advertisement

Technical
Publication
Direction 5461425-8EN
Revision 6
GE Healthcare
Brivo XR118 Service Manual
Copyright © 2015-2019 by General Electric Company, Inc.
GE Healthcare
gehealthcare.com

Advertisement

Table of Contents
loading

Summary of Contents for GE Healthcare Brivo XR118

  • Page 1 GE Healthcare gehealthcare.com Technical Publication Direction 5461425-8EN Revision 6 GE Healthcare Brivo XR118 Service Manual Copyright © 2015-2019 by General Electric Company, Inc.
  • Page 2 GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL Published by the GE Company, Inc. Page 2...
  • Page 3 GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL LEGAL NOTES TRADEMARKS Adobe, the Adobe logo, Acrobat, the Acrobat logo, Exchange, and PostScript are trademarks of Adobe Systems Incorporated or its subsidiaries and may be registered in certain jurisdictions.
  • Page 4 GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL Page 4 Legal Notes...
  • Page 5 GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL LANGUAGE • Deze onderhoudshandleiding is enkel in het Engels verkrijgbaar. ПРЕДУПРЕЖДЕНИЕ Това упътване за работа е налично само на английски език. • • Ако доставчикът на услугата на клиента изиска друг език, задължение...
  • Page 6 • Versuchen Sie nicht diese Anlage zu warten, ohne diese Serviceanleitung gelesen und verstanden zu haben. • Wird diese Warnung nicht beachtet, so kann es zu Verletzungen des Kundendiensttechnikers, des Bedieners oder des Patienten durch stromschläge, Mechanische oder Sonstige gefahren kommen. Page 6 Important Precautions...
  • Page 7 • Ezen karbantartási kézikönyv kizárólag angol nyelven érhető el. • Ha a vevő szolgáltatója angoltól eltérő nyelvre tart igényt, akkor a vevő (HU) felelőssége a fordítás elkészíttetése. • Ne próbálja elkezdeni használni a berendezést, amíg a karbantartási kézikönyvben leírtakat nem értelmezték.
  • Page 8 GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL 경고 본 서비스 지침서는 영어로만 이용하실 수 있습니다 . • 고객의 서비스 제공자가 영어 이외의 언어를 요구할 경우 , 번역 서비스를 • (KO) 제공하는 것은 고객의 책임입니다 .
  • Page 9 GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL ATENÇÃO • Este manual de assistência técnica só se encontra disponível em inglês. • Se qualquer outro serviço de assistência técnica, que não a gems, solicitar (PT-BR estes manuais noutro idioma, é da responsabilidade do cliente fornecer os serviços de tradução.
  • Page 10 GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL VARNING • Den här servicehandboken finns bara tillgänglig på engelska. • Om en kunds servicetekniker har behov av ett annat språk än engelska (SV) ansvarar kunden för att tillhandahålla översättningstjänster.
  • Page 11 In performing all electrical work on these products, GE will use its own specially trained field engineers. All of GE’s electrical work on these products will comply with the requirements of the applicable electrical codes.
  • Page 12 GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL GE personnel, please use the GEHC CQA Process to report all omissions, errors, and defects in this publication. Page 12 Important Precautions...
  • Page 13 GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL Revision History Revision Date Reason for change Dec. 1, 2015 First release Mar 10, 2016 Add explanations for calibration procedures. June. 12, 2016 Add RIS/HIS trouble shooting section.
  • Page 14 GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL List of Effected Pages PAGES REVISION PAGES REVISION 1 to 196 1 to 196 1 to 250 1 to 312 1 to 318 1 to 318 Page 14...
  • Page 15: Publication Conventions

    GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL Preface Publication Conventions Standardized conventions for representing information is a uniform way of communicating information to a reader in a consistent manner. Conventions are used so that the reader can easily recognize the actions or decisions that must be made.
  • Page 16 GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL NOTICE - Indicates information or a company policy that relates directly or indirectly to the safety of personnel or protection of property. This signal word is associated directly with a hazard or hazardous situation and is used in place of 'DANGER,' 'WARNING,' or 'CAUTION.'...
  • Page 17 GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL Section 2.0 Publication Conventions General Paragraph and Character Styles Prefixes are used to highlight important non-safety related information. Paragraph prefixes (such as Purpose, Example, Comment or Note) are used to identify important but non-safety related information.
  • Page 18 GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL Computer Screen Output/Input Text Character Styles Within this publication, mono-spaced character styles (fonts) are used to indicate computer text that’s either screen input and output. Mono-spaced fonts, such courier, are used to indicated text direction.
  • Page 19: Table Of Contents

    Chapter 3 Installation Guide ....................33 Section 1.0 Steps for system installation ....................... 33 Section 2.0 Installation Handoff Checklist ..................... 35 Chapter 4 Target System Check..................... 39 Chapter 5 Installation & Configuration .................. 43 Published by the GE Company, Inc. Page 19...
  • Page 20 Rating Plates ..........................125 Section 14.0 Installation Checklist ........................126 Chapter 6 Calibration ......................129 Section 1.0 Before You Begin........................129 Section 2.0 Calibration Sequence ......................... 131 Section 3.0 Invoking New Calibrations ......................131 Published by the GE Company, Inc. Page 20...
  • Page 21 Regulatory Test .......................... 156 Chapter 8 Troubleshooting..................... 159 Section 1.0 Hardware Failure ........................159 Section 2.0 Software Issues .......................... 172 Section 3.0 Wireless Architecture........................206 Section 4.0 Manual Bad Pixel Calibration...................... 207 Published by the GE Company, Inc. Page 21...
  • Page 22 Digital Detector ........................... 235 Chapter 11 Master Interconnect System Map ............... 239 Section 1.0 Scope............................239 Section 2.0 Symbols/Conventions ......................... 239 Section 3.0 Digital System MIS Map ......................240 Chapter 12 Set Preferences ....................241 Published by the GE Company, Inc. Page 22...
  • Page 23: Chapter 1 Safety And Service

    GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL Chapter 1 - Safety and Service Section 1.0 Lock-Out & Tag-Out (LOTO) System Power Name of Equipment: System Number Of Locks: One per person, working on the system...
  • Page 24 GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL LOTO Procedure for System Power Name of Equipment: System Step Lockout/Tagout Procedure Prepare for shutdown of equipment by notifying affected personnel working in the area that lockout/tagout is being performed.
  • Page 25: Applying System Power

    GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL Section 2.0 Applying System Power System Preparation for Energizing Power POTENTIAL PERFORM LOTO BEFORE PERFORMING SERVICE OF ANY KIND. See Lock-Out & Tag-Out ELECTRICAL (LOTO) on page 23.
  • Page 26: De-Energizing System

    GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL Section 3.0 De-Energizing System Shut down is controlled by its own power socket. There is no recommended shutdown sequence for XR118 system and target system. Complete shutdown takes approximately 80 Seconds.
  • Page 27: Equipment Safety - Electrostatic Discharge (Esd)

    GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL Section 4.0 Equipment Safety - Electrostatic Discharge (ESD) A sudden discharge of static electricity from your finger or other conductor can destroy static- sensitive devices or microcircuitry. Often the spark is neither felt nor heard, but damage occurs. An electronic device exposed to electrostatic discharge (ESD) might not appear to be affected at all and can work perfectly throughout a normal cycle.
  • Page 28 GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL • Heel straps, toe straps, and boot straps can be used at standing workstations and are compatible with most types of shoes or boots. On conductive floors or dissipative floor mats, use them on both feet with a maximum of one-megohm +/- 10% resistance between the operator and ground.
  • Page 29 GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL 4.5.1 Personnel Requirements Personnel Requirements Preliminary Reqs Procedure Finalization Not Applicable 30min 10min 4.5.2 Overview This procedure can be performed with: • Fluke tester 4.5.3 Tools and Test Equipment Item Qty.
  • Page 30 GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL • Power on the Tool - Press the power switch on the left-side panel so the “I” side of the ac power switch is depressed. The tool will perform a series of self tests and then display the message shown as below picture when the self test has completed successfully.
  • Page 31 GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL 2. Disconnect the power cord from the site receptacle and disconnect the PC adapter output from the PC terminal. 3. Connected the alligator clip of the black test lead to power cord PE pin. Connected the alligator clip of the red test lead to test point.
  • Page 32 GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL Page 32 Section 4.0 - Equipment Safety - Electrostatic Discharge (ESD)
  • Page 33: Chapter 2 Before You Begin

    GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL Chapter 2 - Before You Begin Section 1.0 Introduction The goal of this document is to guide the service representative through an entire X-ray room installation. The following On-site Requirements must be accounted for prior to beginning installation of the system.
  • Page 34 GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL Interconnection Interconnection of products by furnished cables or contractor supplied wiring is indicated by numbered MIS runs in the installation sections. Run interconnect charts are available, which describe the connection and function of each wire.
  • Page 35: On-Site Requirements

    GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL Section 3.0 On-Site Requirements Tools and Test Equipment There are no additional tools and test equipment besides the standard service representative tool kit which must be available before installing the Brivo XR118 System as below.
  • Page 36: Installation Process

    GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL The Brivo XR118 System for fixed RAD is divided in to basic components: 1.) Digital Detector 2.) Workstation 3.) Dock 4.) 19" LCD Monitor 5.) Access Point 6.) Power Input & Standard Adapter 7.) Bar Coder...
  • Page 37: Chapter 3 Installation Guide

    GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL Chapter 3 - Installation Guide The primary tasks are designed to be performed in sequence. Follow the sequence of tasks described for each primary task in Chapter 3 - that follow.
  • Page 38 GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL AEC (Optional) Leakage Tests People Required: 1 Minutes to Complete: 15 Leakage Tests Labels and Rating Plates People Required: 1 Minutes to Complete: 5 Attach labels and rating plates...
  • Page 39: Installation Handoff Checklist

    GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL Section 2.0 Installation Handoff Checklist Pre-Installation Checklist Delivery Date: Sales Person: Customer: FDO No.: Room # Equipment: Physical Requirements of Site Completed 1.) Room size adequate for intended equipment configuration? 2.) Floor is strong enough for intended equipment and mounting methods...
  • Page 40 10.) Room IP Addresses for DICOM and Broadband identified? 11.) Broadband connection provided for InSite connection? 12.) Have all fire/safety inspections for occupancy been completed? 13.) Send completed checklist to the GE Healthcare installation team. Comments: Inspection Date(s): Installation Project Manager Signature Page 40 Section 2.0 - Installation Handoff Checklist...
  • Page 41 GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL Installation Checklist Purpose: Installation checklist to confirm that all critical installation steps have been completed.. Item # Item Sign Off Power and ground cables connected: Power cables (how many?, where?) All signal cables connected and tightened.
  • Page 42 GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL Item # Item Sign Off 27.) Has backup of original system been performed? (Patient Data Base as well as total system backup) 28.) Has Application Specialist received all checklists:...
  • Page 43: Chapter 4 Target System Check

    GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL Chapter 4 - Target System Check Digital X-Ray Imaging System (Brivo XR118) can upgrade target analog system which meets the below requirements to DR system. Before installation XR118, purchaser needs to check target system from two sides.
  • Page 44 GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL Table 4-1 Target System Requirements Categor Item Preferred Capability Target Pass System /Fail Claim System Regulatory Shall comply with local medical equipment regulatory requirement, such as GB9706/ IECv2/IECv3...
  • Page 45 GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL Target System Grid Type Recommend Grid Type for GLR 60 LP/CM 60LP/CM 62 LP/CM 36LP/CM, 60LP/CM 64 LP/CM 36LP/CM 66 LP/CM 36LP/CM 68 LP/CM 70LP/CM 70 LP/CM 70LP/CM...
  • Page 46 GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL Page 46...
  • Page 47: Chapter 5 Installation & Configuration

    GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL Chapter 5 - Installation & Configuration Section 1.0 Before You Begin Before installation work begins, the site should be checked to verify the correctness of junction box sizes, their locations, the number and sizes of interconnecting wiring and proper power availability in the room.
  • Page 48: Cables

    GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL Section 2.0 Cables Note: All system cable connection, please refer to the marks on the cable. Routing • System MIS cables should be pulled to/from the System Components before the System Components is moved into the final position.
  • Page 49: Unpacking Main Box

    GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL Section 3.0 Unpacking Main Box When moving the system package or components, two-man lift is required. Be careful not exposure to the sharp edge of broken components. Figure 5-3 Components for Fixed RAD system Components list for Fixed RAD system 1.) Image Workstation...
  • Page 50 GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL Open pink box (OPEN ME FIRST box) at first. Take out the Manual CD. Read Manuals carefully before system installation and calibration. Figure 5-4 Open Me First Page 50...
  • Page 51 GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL Recommended Control Room Placement Position the components according to the site plan. 3.1.1 Unpacking Components Boxes Image Workstation Dock Figure 5-5 Unpacking Components Boxes 3.1.2 Room Layout It’s recommended to place the XR118 system near to the target system console Figure 5-6 Recommended Placement Chapter 5 - - Installation &...
  • Page 52: Mis Cable Connections

    12/Aug/2014 Brivo XR118 FIRST MADE FOR: APPR Yang Yang 02/Dec/2015 Only support fix RAD upgrade GE HEALTHCARE (TIANJIN) 6118888SCH SHT/CONT ON OTHER COMPANY LIMITED Figure 5-7 MIS Map for Fixed RAD System Page 52 Section 4.0 - MIS Cable Connections...
  • Page 53 GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL 1.) Connect image workstation embedded battery. Screwdriver T10 Demount the battery cover Remove the cover Connect the battery connector and Battery connector Remount the cover 2.) Connect switching power supply with power plug.
  • Page 54 GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL 3.) Connect switching power supply to Dock’s DC power port. 4.) Connect Bar-coder, USB DVD driver to Dock’s USB ports. 5.) Connect the Keyboard, Mouse to Dock's USB port.
  • Page 55 Ethernet PoE port shall be only used to connect wireless AP. 8.) Insert the image workstation into dock’s charging slot. Note: Insert the workstation with the GE side facing the front side of dock to ensure proper orientation. 9.) Connect monitor's power cable to Hospital main power.
  • Page 56: Ap Assembly Preparation

    GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL Section 5.0 AP Assembly Preparation Note: Because wireless performance will be tested in AP Assembly Installation on page 122 and AP Assembly location may be changed, please do not install AP Assembly in this step, just place it at Default location or Adjusting range according to room layout.
  • Page 57 GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL Assembly Placement Default Location Adjusting Range Detector Insert Direction Default Location 3 < 4 meters Table Exam Room Default Location 1-1 Default Location 1-2 (Recommended) (Recommended) * Default Location 1-1: Only table bucky is required.
  • Page 58: Detector Installation

    GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL Section 6.0 Detector Installation Unpacking There are two locks on each side of detector package. Figure 5-10 Detector Package Shell Notch Knob Alignment Rotate the knob to align with notches of the shell...
  • Page 59 GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL Figure 5-12 Open the box Detector Battery Installation To replace a battery: 1.) Release the two latches at the back of detector. The battery pops up for replacement.
  • Page 60 GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL Detector Placement Note: More detailed information about detector placement, please refer to the “Digital Detector Application” chapter in Operator Manual. 6.4.1 Table upgrade 1.) Remove the film cassette from the table housing if any.
  • Page 61: System Power Up

    GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL Section 7.0 System Power Up 1.) Plug the Switching Power Supply and monitor power cable into hospital main power (If required). 2.) The dock and AP automatically boot up.
  • Page 62: Software Installation

    GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL Section 8.0 Software Installation Before You Begin Potential for Data Loss. Archive all protocols, patient files, and backup calibration and configuration data. It's recommended that a system backup be performed prior to any installation, upgrade or restoration of software.
  • Page 63 GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL The following data is not included in a system backup: • Archival of Image Data - The process for archiving image data is found in the operators manual.
  • Page 64 GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL 8.2.3 Backup Protocol Data Base Note: From SUIF/Configuration/General/Application Language, record the language config. This is because the database only can be restored at the system configured with the same language environment.
  • Page 65 GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL 8.2.4 Backup Images Note: Pressing the eject button on the DVD drive will not open the CD tray. While working with the Image Management screen open, always use the eject option within the [Tools] drop-down list.
  • Page 66 GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL System Software Installation 8.3.1 Personnel Requirements Required Preliminary Finalizati Procedure Persons Reqs 15 (Time estimate does not include saving 5 mins and restoring patient images and/or 0 min protocols files) 8.3.2...
  • Page 67 GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL 8.3.3 Preliminary Requirements 8.3.3.1 Tools and Test Equipment Effectivit Item Part# Manufacturer Linux System Software DVD (5589170) 2 Detector Switch Software Patch CD(5589170-1) 8.3.3.2 Consumable Materials Effectivit Item...
  • Page 68 GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL and press the <Enter> key, then the installation starts. Figure 5-13 System Installation - Step 1 iv.) Be patient, it takes approximately 15 minutes for the system software to load.
  • Page 69 GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL Note: Please wait this page to disappear.. vii.) Remove the software DVD from the USB DVD drive. 6.) Login to System Note: Username and password are case sensitive. Enter them carefully.
  • Page 70 GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL sudo su mount -o loop /dev/sr0/ /mnt/cdrom/ cd /mnt/cdrom ./InstallPatch.sh 4.) Wait for message – “Note: atlasidc_step9_patch_release-2017_FW36.5-RC1-1.x86_64.rpm patch installation completed. Please restart system.” 5.) Then perform below command:...
  • Page 71 GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL Restore 8.4.1 Before You Begin Note: The detector calibration data table_IDC can not be restored in different system. Note: If you wish to use a USB flash drive, select [USB DRIVE] and install drive into USB port.
  • Page 72 GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL 7.) Click on the down-arrow in the drop down box to display a drop down list of items. 8.) Click on the item you wish to restore. 9.) Click on [Continue].
  • Page 73 GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL 8.4.3 Restore Protocol Data Base 1.) Be assure the system language config are the same with the backed database. From the Worklist screen, click the Utilities icon. The Utilities screen appears.
  • Page 74: System Configuration

    9.1.3 Addition of Purchased Options Installation of purchased options can only be performed by your authorized GE sales and service representative. If you add purchased options to your system, you should update the system configuration files immediately and create and save a new backup.
  • Page 75 9.3.2 Site Parameters The following information can be viewed regardless of service mode. All can be edited, except for those identified by an asterisk “*” (which require setup by GE personnel). Table 12-2 Site Parameters Parameter Definition Name of specific location (Hospital, clinic...) where system is installed (64...
  • Page 76 RIVO ERVICE ANUAL Parameter Definition Service Contract Customer licensed Inhouse Service Contract number; created by GE Number* Healthcare. (64 characters maximum) Licensed Inhouse Service Contract number expiration date; created by Contract Expiration* GE Healthcare. (64 characters maximum) 16 characters maximum...
  • Page 77 GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL Hospital Network Configuration Go to SUIF -> Configuration -> Networking. After selected and committed the change, reset system. Table 12-4 Wired Fixed IP Settings Parameter Definition If enabled (Yes), the hospital wired IP interface will acquire IP settings from DHCP server on the hospital network.
  • Page 78 GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL Hardware AP Configuration For the countries where wireless communications applications are not allowed by the regulations, the wireless image detector shall be used after selecting the wireless regulatory code during configuration in SUIF.
  • Page 79 • Only viewable by GE service personnel. • Configure source of InSite connectivity. • Default values only allow communication from the GE back office. • Could require configuration for custom environments (Refer to the remote service connectivity team.) • Additional “allowed” nodes not viewable by the user.
  • Page 80 EVISION RIVO ERVICE ANUAL • Insite & ILinq checkout could be successful with PNF On. Please select GE Service office as Standard(150.2.0.0/16) and Access Level as Pre-defined setting for encrypted access. 9.6.2.2 Buttons 1.) Firewall (Top Box) • Turns firewall on and off.
  • Page 81 GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL 1.) Select Product Network Filter from the Configuration menu. 2.) Select the [Expert] Tab. 3.) For TVA/RCOC support, do the following: a.) Port 5800 I.) Enter the Port number 5800.
  • Page 82 GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL User Accounts 9.7.1 System Log-in Upon system start-up, you must enter an username and password to enable the full functionality of the system. When the log in pop-up window is displayed, use the keyboard to enter your username and password.
  • Page 83 GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL Operator Manual (OM) Installation The operator manual must be installed: when the system is initially installed following a complete system load from cold following a software reload of the workstation PC The procedure follows: 1.) Verify that no archival process is running.
  • Page 84 GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL DICOM Configuration See the Operator’s Manual for instructions on how to configure DICOM on this system. The following worksheets should be used to capture and archive your system’s DICOM configuration for future reference.
  • Page 85 GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL • Option 5 - Checking this option will cause the workstation PC to strip out all patient information in the images being sent to the remote node being configured. Same as de-identify and push automatically.
  • Page 86 GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL 9.10 DICOM Printers Note: For DICOM printer, please acquire mandatory parameters (Application Entry Title \ Network Address \ Port Number) from hospital supervisor. For other parameters, please configure exactly according to the DICOM printer data sheet.
  • Page 87 GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL Attribute Value Magnification Alternative Print Mode Alternative Magnification Orientation Media Type Destination Media Size Format Chapter 5 - - Installation & Configuration Page 87...
  • Page 88 GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL 9.11 Some Recommended Printer Configurations Note: For Brivo XR118, only integer value for the Printer Pixel Size is accepted. Please round this value to the nearest integer. 9.11.1...
  • Page 89 GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL Attribute Value Comments Film 8 x 10in Pixel Width = 2388 Pixel Height = 2972 AGFA Layout - Portrait Sizes 9006 10 x 12in Pixel Width = Pixel Height =...
  • Page 90 GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL FILM SIZE FORMAT AGFA LAYOUT 11 x 14 9014 11 x 14 9102 11 x 14 9103 11 x 14 9104 11 x 14 9018 11 x 14...
  • Page 91 GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL 9.11.2 Agfa Drystar 5500 Printer Parameters Table 12-15 Agfa Drystar 5500 Printer Parameters Attribute Value Comments Printer DICOM Printer Label DRYSTAR5500 Or customers preference Application Entry Title Enter value given by Site Network Administrator...
  • Page 92 GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL Attribute Value Comments Film 8 x 10in Pixel Width = 3852 Pixel Height = 4300 AGFA Layout - Portrait Size 9006 10 x 12in Pixel Width = 4880...
  • Page 93 GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL FILM SIZE FORMAT AGFA LAYOUT 11 x 14 9204 8 x 10 9006 8 x 10 9009 8 x 10 9008 8 x 10 9010 8 x 10...
  • Page 94 GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL 9.11.3 Fuji FM DPL Print Parameters Table 12-17 Fuji FM DPL Print Parameters Attribute Value Comments Printer DICOM Printer Label FUJIFMDPL Or customers preference Application Entry Title Enter value given by Site Network Administrator...
  • Page 95 GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL Attribute Value Comments The following configurations must be configured by the Fuji printer representative to print correctly sized images: • LUT1 as SAR17 for calling AE_TITLE. • Trim width at 1 pixel for calling AE_TITLE.
  • Page 96 GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL 9.11.4 Fuji DryPix 1000 Print Parameters Table 12-18 Fuji DryPix 1000 Print Parameters Attribute Value Comments Printer DICOM Printer Label DRYPIX1000 Or customers preference Application Entry Title Enter value given by Site Network Administrator...
  • Page 97 GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL Attribute Value Comments Layout Select the desired layouts 1:1, 1:2, 2:1, 2:2 allowed for this printer Slide formats Not Applicable Film 8 x 10in Pixel Width = 2280...
  • Page 98 GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL 9.11.5 Fuji DryPix 3000 Print Parameters Table 12-19 Fuji DryPix 3000 Print Parameters Attribute Value Comments Printer DICOM Printer Label DRYPIX3000 Or customers preference Application Entry Title Enter value given by Site Network Administrator...
  • Page 99 GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL Attribute Value Comments Film 8 x 10in Pixel Width = Pixel Height = Size 10 x 12in Pixel Width = Pixel Height = 10 x 14in Pixel Width =...
  • Page 100 GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL 9.11.6 Fuji DryPix 7000 Printer Parameters Table 12-20 Fuji DryPix 7000 Printer Parameters Attribute Value Comments Printer DICOM Printer Label DRYPIX7000 Or customers preference Application Entry Title Enter value given by Site Network Administrator...
  • Page 101 GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL Attribute Value Comments Film 8 x 10in Pixel Width = 3907 Pixel Height =4819 Size 10 x 12in Pixel Width = Pixel Height = 10 x 14in Pixel Width = 5075...
  • Page 102 GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL 9.11.7 Kodak 8200 Print Parameters Table 12-21 Kodak 8200 Print Parameters Attribute Value Comments Printer DICOM Printer Label KODAK8200 Or customers preference Application Entry Title Enter value given by Site Network Administrator...
  • Page 103 GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL Attribute Value Comments Film 8 x 10in Pixel Width = Pixel Height = Size 10 x 12in Pixel Width = Pixel Height = 10 x 14in Pixel Width =...
  • Page 104 GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL 9.11.8 Kodak 8500 Print Parameters Table 12-22 Kodak 8500 Print Parameters Attribute Value Comments Printer DICOM Printer Label KODAK8500 Or customers preference Application Entry Title Enter value given by Site Network Administrator...
  • Page 105 GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL Attribute Value Comments Film 8 x 10in Pixel Width = Pixel Height = Size 10 x 12in Pixel Width = Pixel Height = 10 x 14in Pixel Width =...
  • Page 106 GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL 9.11.9 Kodak 8700 Print Parameters Table 12-23 Kodak 8700 Print Parameters Attribute Value Comments Printer DICOM Printer Label KODAK8500 Or customers preference Application Entry Title Enter value given by Site Network Administrator...
  • Page 107 GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL Attribute Value Comments Film 8 x 10in Pixel Width = Pixel Height = Size 10 x 12in Pixel Width = Pixel Height = 10 x 14in Pixel Width =...
  • Page 108 GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL 9.11.10 Kodak 8900 Print Parameters Table 12-24 Kodak 8900 Print Parameters Attribute Value Comments Printer DICOM Printer Label KODAK8900 Or customers preference Application Entry Title Enter value given by Site Network Administrator...
  • Page 109 GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL Attribute Value Comments Film 8 x 10in Pixel Width =4742 Pixel Height =6286 Size 10 x 12in Pixel Width = 6286 Pixel Height = 7342 10 x 14in...
  • Page 110 GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL 9.11.11 Codonics Horizon Print Parameters Table 12-25 Codonics Horizon Print Parameters Attribute Value Comments Printer DICOM Printer Label HORIZON Or customers preference Application Entry Title Enter value given by Site Network...
  • Page 111 GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL Attribute Value Comments Film 8 x 10in Pixel Width = 2406 Pixel Height = 2790 Size 10 x 12in Pixel Width = 6286 Pixel Height = 7342 10 x 14in...
  • Page 112 GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL 9.12 Adding a Bar Code Reader 9.12.1 Configuration The following sections provide instructions on the most common configuration choices. A User’s Guide is supplied with the bar code reader. Please refer to the User’s Guide for additional programming choices.
  • Page 113 GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL 9.12.4 Trigger The following bar codes will allow you to use the reader in Manual Trigger mode (need to press the trigger to read), Automatic Trigger mode (the beam is always on), or Presentation mode (the reader is activated when it “sees”...
  • Page 114 GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL 9.12.8 Data Format Editor Instructions The following are abbreviated instructions for entering a data format. 1.) Scan the Enter Data Format symbol. 2.) Primary/Alternate Format: scan 0 for Primary Format 3.) Terminal Type: scan 099 for any terminal type...
  • Page 115 GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL 9.13 Configure Chinese Input Bar and Change Its Font Size 1.) Go to SUIF -> Configuration -> General. Select “Encoding” to “GB1830”, “Application Language” to “Chinese”, and then click the “Commit” button. After the configuration, restart the system.
  • Page 116 GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL 4.) Select Panel->GTK-> font from the setup UI 5.) Select the appropriate font family style, especially the size and press “OK” button. The new font size for the Chinese input tool will be applied.
  • Page 117: Detector Association

    GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL Section 10.0 Detector Association 10.1 Detector Firmware Download Go to XR118 SUIF / Utilities/Subsystem Software/Det Data Download, download detector firmware. Note: This step must be performed once a new detector is used for this system.
  • Page 118 GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL 10.2 Detector Registration A detector must be registered with an x-ray system at the first time when the two work together. The registration is accomplished by connecting the detector and the 118 system through the registration process.
  • Page 119 GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL 4.) Enter the new Detector Name. 5.) Chose to make the new Detector Associated or Non-Associated. 6.) Press [Save]. The new Digital Detector is now registered with the system.
  • Page 120 GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL 10.3 Wireless Detector Management Click the Detector Management icon to list all the detectors registered in the system. Figure 5-18 Detector Management To associate a “Non-associated”: 1.) Selects a non-associated detector from the detector list which has been boot up.
  • Page 121 GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL To release an associated detector 1.) Selects the associated detector from the detector list which has been boot up. • When the associated detector is selected, the “Associate” button changes to “Release”.
  • Page 122: Ap Assembly Installation

    GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL Section 11.0 AP Assembly Installation 11.1 AP Assembly location selection 11.1.1 AP location selection process 1.) Place the AP at Default location and then perform wireless signal strength test. If passed, install AP at default location.
  • Page 123 GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL 11.1.2 Wireless Signal Strength Test Go to SUIF -> Diagnostics -> Wireless Detector -> Signal Strength. Click the “Start” to test the wireless communication quality. If there is no bucky on the wall stand, just test the wireless signal with detector in the table bucky.
  • Page 124 GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL 11.2 AP Assembly Installation Note: Before installing AP Assembly, make sure the proper installation location has been determined. There are two ways to mount or place the AP: Wall mounting and Table placement...
  • Page 125 GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL 2.) Drill the four 40mm deep holes with 6mm diameter masonry drill bit, and then install the wall anchors. Hammer the anchors Drill holes Leave 5mm distance above the wall Insert the Screws 3.) Mount the wall adapter, then fix the screws with wall adapter, make sure the wall adapter is...
  • Page 126 GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL 4.) Mount the AP with wall adapter. 5.) Connect the ethernet cable and USB cable with AP. 11.2.2 Shelf/Table placement (not anchored) Note: Only consider this method when the wall mounting is not permitted, for example lead shielding might be damaged.
  • Page 127: Leakage Current Test

    GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL Section 12.0 Leakage Current Test 12.1 Leakage Current Measurements LEAKAGE CURRENT MEASUREMENTS When a system installation is completed, it must be inspected for leakage current levels by qualified personnel.
  • Page 128 GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL 12.1.1 Personnel Requirements Personnel Requirements Preliminary Reqs Procedure Finalization Not Applicable 30min 10min 12.1.2 Overview This procedure can be performed with: • Fluke tester 12.1.3 Tools and Test Equipment Item Qty.
  • Page 129 GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL 2) Refer to 4.5.5 Set up tools on page 29 Setting the test standard, select the test standard to IEC60601-1. 3) Press the “uA” button to access Leakage Current Test interface.
  • Page 130 GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL 4) Press the “POLARITY” button to “NORMAL”, press the “NEUTRAL” button to “CLOSED”, then power on the system and record the earth leakage current value as below. 5) After earth leakage current test, shut down the system and reconnect the power cord to the site receptacle, then power on the system.
  • Page 131: Rating Plates

    GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL Section 13.0 Rating Plates Attach multilingual labels to overlay the Chinese labels attached in factory. 13.0.1 System Rating Plate Label Description References Digital X-Ray Imaging System 5501790 rating plate Chapter 5 - - Installation &...
  • Page 132: Installation Checklist

    GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL Section 14.0 Installation Checklist Purpose: Installation checklist to confirm that all critical installation steps have been completed.. Item # Item Sign Off Power and ground cables connected and tightened: Power cables (how many?, where?) All signal cables connected and tightened.
  • Page 133 GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL Item # Item Sign Off Has system documentation (paper and CD) been given to the customer? 22.) Also, for VA sites, has 2nd set of documentation been given to the customer? Is the hospital scheduling system (HIS/RIS) configured? Does the modality display MWL 23.)
  • Page 134 GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL Page 134 Section 14.0 - Installation Checklist...
  • Page 135: Chapter 6 Calibration

    GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL Chapter 6 - Calibration Section 1.0 Before You Begin Service User Interface (SUIF) A list of available system calibrations is displayed on the Service User Interface, whenever the calibration tab is chosen.
  • Page 136 GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL The use of Collimator tape (Note the difference between SID and OID) In the calibration, you need to use collimator tape to measure SID. In the case to measure SID, you...
  • Page 137: Calibration Sequence

    GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL Section 2.0 Calibration Sequence Potential for Data Loss 1.) Always create an up-to-date backup of your system data following product installation or after making any calibrations. 2.) The system must be reset, upon exiting the Service User Interface for Configuration and Calibration information to take affect.
  • Page 138: Detector Calibration

    GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL Section 5.0 Detector Calibration Before You Begin The following applies when performing Detector (Gain, Bad Pixels) calibrations. 1.) Follow the instructions presented on SUIF. 2.) Correct detector firmware must have been downloaded to the detector before proceeding.
  • Page 139 GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL • When a pixel or line artifact appears in an image(s). Gain Calibration Procedure There's the potential for data loss. Be aware of the following: • Data loss can occur if procedures are not followed in the proper sequence.
  • Page 140 GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL You are required to do the following: 2 triangle silk-prints on detector Table mode Wallstand mode 1.) In Detector calibration/QAP process, system will evaluate Quality Check. To save time and make QC check pass, firstly make exposure to take an image before calibration with SID=150cm/80KV/50mA/5mAs, use ring tool to check if 4 corners’...
  • Page 141 GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL 8.) Execute Gain Calibration refer to the screen instruction. Note: If QC check failed, Do not continue and click the cancel button to redo QAP based on instructions on the QC check failed screen.
  • Page 142 GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL Bad line example: After manual BP, take an extra exposure to confirm the specified bad line disappear. Detector Replacement If detector has too many bad pixels, redo calibration again, if still fail, suggest to replace detector.
  • Page 143: Detector Qap

    GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL Section 6.0 Detector QAP Before You Begin The following applies when performing Quality Assurance Procedure (QAP). • Follow the instructions presented on SUIF. When to Perform QAP On a scheduled, weekly basis or QAP required icon is shown.
  • Page 144: Emi Reduction

    GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL Section 7.0 EMI Reduction EMI Check Procedure If the RCN in QAP always failed or row correlated noise (RCN) type of artifacts found in image, following test need to be done to double check the EMI.
  • Page 145 GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL EMI Reduction In order to allow modification of the behavior of the EMI Reduction algorithm, several changeable parameters will be present on the Service User Interface (SUIF). These parameters are: amplitude threshold, minimum correctable frequency, maximum correctable frequency, maximum allowable correction and maximum number of iterations.
  • Page 146 GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL 3.) Maximum Allowable Correction The EMI Reduction algorithm has a maximum allowable correction. Any correction [amplitude since the correction removes sinusoids] to the image will be smaller than or equal to this value.
  • Page 147: Monitor Setting

    GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL Section 8.0 Monitor Setting Before You Begin Before starting Image Monitor Set-Up, verify the followings are completed: System has been powered up and monitor has been on for at least 20 minutes.
  • Page 148: Printer Calibration

    GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL Section 9.0 Printer Calibration The following applies when performing Printer calibrations. Note: Only performing printer calibration when printer has problem. Follow the instructions presented on SUIF. Section 10.0 PACS / Image Display Test Image 10.1...
  • Page 149 GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL a.) Load Images i.) Load the PACS Test Images onto the acquisition workstation. Either use: • DICOM-formatted CDROM with images • Images on system under Image Tools ii.) Patient Name will appear as GE_PACS_Test_Image, Definium iii.) Patient ID will be 1000001...
  • Page 150 Review images on PACS i.) Bring PACS Test Images 1 and 6 up on the PACS review workstation. • The images can be identified by the watermark image number next to the GE logo inside the chest image region. •...
  • Page 151 GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL GE-IDX-RealTime iPACS Viewer v5.0 Picker Philips EasyAccess Neusoft v3.0 Sectra v10.2 Thinking Systems - - - - - AMICAS LightView v4.10.196 10.3 Results Table 6-2 Results from the Simplified and Additional Tests Sections...
  • Page 152 Each PACS Test Image consists of two aspects: • A clinical chest image on the right side with Image Number and a GE monogram watermark • A series of twelve vertical bands on the left side (Top six Dark Bands and Bottom six Bright...
  • Page 153 GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL PACS / Image Display Test Image The 12 bands on the left side each have a different brightness. The top six should appear “dark” and the bottom six “bright”.
  • Page 154 Use the values in the following table to quantitatively compare images. For example, to determine if a PACS supports GE VOI-LUTs, compare PACS Test Images 1 and 6. If the number of distinct bands and visible squares are equal, then the PACS correctly supports VOI-LUTs.
  • Page 155 GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL 10.5 Additional Tests The following tests are meant to provide additional information about PACS behavior and performance. These tests should be used if further troubleshooting is required beyond the configuration determined in the “Simplified Procedure”.
  • Page 156 GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL on PACS through either a menu drop down, a button, or right-clicking the image. Click something indicating “ information ”, “ i ”, or “ header ”. Contact local IT support for further information.
  • Page 157 GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL 3.) Visible WC and WW Values a.) No-Burn Mode (PACS Test Image 1) i.) The DICOM header of each PACS Test Image is always sent with WC/ WW=5226/2654.
  • Page 158 GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL a.) Open the PACS Test Image 6. b.) Change the Window-Level on PACS trying to see the 5 squares in all 12 bands: A suggested value is WC/WW=0/500 (Note: this may vary between different PACS workstations) The loss of dynamic range is most common in the top dark band.
  • Page 159 8.) Multiple VOI-LUTs As indicated above, the GE acquisition workstation is capable of sending 3 VOI-LUTs in the DICOM header. Some PACS recognize these 3 VOI-LUTs and allow the user to change between each VOI-LUT. The user can change VOI-LUTs on PACS by selecting Normal, Harder, or Softer options.
  • Page 160 GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL a.) Configure the PACS to reject the DX-modality b.) Send the PACS Test Image 9 Open the PACS Test Image 9 d.) Open the DICOM header for Image#9 •...
  • Page 161: Chapter 7 Brivo Xr118 System Check

    GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL Chapter 7 - Brivo XR118 System Check Section 1.0 Image Quality Test 1.) Go to SUIF according to the instruction described above. Then press the “Image Quality” button on the SUIF.
  • Page 162: Acquisition Test

    GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL 7.) Set exposure techniques as: SID 150cm; 60kV; 5mAs. 8.) Take exposure 5 times for each exposure mode (Detector under the table and above the table). 9.) After exposures done, view Raw Image with cobb tool at same location to compare the average value.
  • Page 163 GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL This page is intentionally left blank. Chapter 7 - - Brivo XR118 System Check Page 163...
  • Page 164 GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL Page 164 Chapter 7 - - Brivo XR118 System Check...
  • Page 165: Chapter 8 Troubleshooting

    GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL Chapter 8 - Troubleshooting Section 1.0 Hardware Failure Using the Built-in Diagnostics tools Under the Diagnostics tab of SUIF, a set of tools are provided to GEHC service personnel, when the GEHC restricted service key is installed.
  • Page 166 GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL Using the Enhanced Error Messages Enhanced error messages are available to GEHC personnel, when the GEHC restricted service key is installed. The enhanced error messages contain additional information not found in the system error log during normal system operation or available to the normal user.
  • Page 167 GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL Wireless Debug 1.3.1 Detector connection lost 1.) Go to SUIF -> Diagnostics -> Wireless Link. 2.) Click “Start” button. 3.) The wireless link diagnostics result and the recommended debug actions will be displayed.
  • Page 168 GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL 1.3.2 Exposure inhibit or long time of image display (> ~25s) 1.) Go to SUIF -> Diagnostics -> Wireless Detector -> SignalStrength. 2.) Click “Start” to begin the test.
  • Page 169 GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL Image Workstation 1.4.1 General Information for Image Workstation Part Components Detailed Info. SUIF How to Access? Please find this info in SUIF Serial Number Actual Max Capacity to Default Design...
  • Page 170 GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL 1.4.2 Workstation stops working after unplugged from dock Unplug workstation, workstation stops working Connect workstation with power Adapter, power on workstation Workstation boots up? Replace mainboard Replace workstation battery 1.4.3...
  • Page 171 GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL 1.4.4 Workstation battery charging issue Battery capacity is not full and indicator not Workstation battery charging issue changes or the charging process is very slow Power adapter is connected...
  • Page 172 GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL 1.4.5 System boots failure (For all configuration) For All Configuration The system boots failure Four short buzzer warnings? Workstation LED1 green flash Workstation LED1 is red? (3 mins later)?
  • Page 173 GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL 1.4.6 Workstation LED1 is red Workstation LED1 is red? Connect to external power source (Power adapter/Dock) LED1 changes to green? System boots up? Unplug and reinstall SSD drive...
  • Page 174 GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL 1.4.7 System boots failure System boots failure System enters into BIOS without interference? Prompts SSD drive failure? The SSD drive exists in the BIOS? Unplug and reinstall SSD drvie...
  • Page 175 GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL 1.4.8 No display on the Monitor No display on the monitor Is there any operation issue? Please operate the system correctly Check the power output of power Dock power LED is light?
  • Page 176 GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL Detector Battery 1.5.1 Detector Battery Cannot be Charged Detector battery cannot be charged Check the dock power is OK? Please connect the dock properly Check the power output of power...
  • Page 177 GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL Bar code Reader The Bar code reader automatically performs self-tests whenever initially turned on. If it’s not functioning properly, review the following sections to try to isolate the problem or replace the reader.
  • Page 178: Software Issues

    GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL Section 2.0 Software Issues Common Issues 1.) How to find the boot failures? System boot has failed, i.e. UI did not come up or buttons are disabled after boot. Following are the steps.
  • Page 179 GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL Step 1: You need to find out when exactly system crash has happened and who crashed. Open /magichome/xruser/log/sysError.log Go to the end of log file. [if you are using vi editor, press shift + g]...
  • Page 180 GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL 5.) How to start Engineering Tool Box? • If there is no monitor, connect to Workstation with VNC. • Make sure iDR workstation is on the application UI, if not, press “Ctrl + Alt + F1” to switch to application UI, and then press “Ctrl + Shift + F5”...
  • Page 181 GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL 1.) How to take snapshot when UI hang up: > Ctrl + Alt+ F2.F6: switch to console window > Log in with “xruser”, “4$xray” > input snapshot name after ran cmd below: cd /magichome/xruser/engg/ /collectLogs.sh...
  • Page 182 GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL The steps to handle this special case are following. Login with the root’s password #superxr Input “fsck –y” to fix the crashed file system automatically. After above operation done, input “reboot” to restart the Workstation.
  • Page 183 GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL DICOM 2.2.1 DICOM Tool (Viewing Header Information w/o PACS Transfer) Problem: Cannot easily view the DICOM header without transferring an image to PACS. Solution: You can use the DICOM tool in the Image Management UI. To access this tool: 1.) On the Image Management screen, select an image.
  • Page 184 GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL 2.2.2 DICOM Association Time Outs Problem: DICOM Connectivity Intermittent issue. Before concluding that the time out values are too small, please confirm that the Network/ System configuration is correct on either end. Time-out values should be increased only if the Network traffic is too slow for an association to be established in the allotted time.
  • Page 185 GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL 2.2.3 Problem With Worklist From RIS/HIS Problem 1: Some of the worklist entries cannot be received from RIS/HIS system. Figure 8-22 Error system message Solution: Worklist entries could be rejected if following restrictions are NOT followed in DICOM data received from HIS/RIS system.
  • Page 186 GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL Tag Name Group Element VR Type Restrictions ScheduledProcedur 0040 0003 1.) Must be present and value could not be empty. eStepStartTime 2.) Length:14 bytes maximum. 3.) Format HHMMSS.FFFFFF, HH contains hours (range "00"...
  • Page 187 GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL Tag Name Group Element VR Type Restrictions Patient's Weight 0010 1030 1.) Must be present but the value could be empty. 2.) Length:16 bytes maximum. 3.) “0”-”9”, “+”, “-”,“E”, “e”, “." of Default Character Repertoire.
  • Page 188 GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL Problem 2: Some of the images acquired cannot be saved. Figure 8-23 Image saving error pop-up message Solution: Image installation and copy exam could fail if following anomalies exist in DICOM data received from HIS/RIS system.
  • Page 189 GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL 2.2.4 How to Find Out Invalid Value of Tag When query from RIS server, iDR system can discard the workflow entries if any entry contains invalid value of tag. These entries will not be shown the on Workflow list. How to find out the...
  • Page 190 GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL (Figure 2) Take below message as example to show you how to interpret the error message in above screen shot. ERROR MESSAGE: [XRWLDataValidator.java:157] Tue, 17 May 2016 00:06:36.76 INFO [ThreadId : 1030] [ERROR]tag name:(10,30),value:1980 01 01,caused by Rejecting the...
  • Page 191 GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL (Figure 3) Then we can update value of tag within the entry in RIS, refer the red box on figure 4. Refreshing workflow entries on iDR, the entry can show on workflow list.
  • Page 192 GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL (Figure 4) Appendix 1 We list some DICOM tags which are commonly used as below, you can refer them when you check error log. (0008,0005) SpecificCharacterSet (0008,0050) AccessionNumber SH 1...
  • Page 193 GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL (0008,1084) AdmittingDiagnosisCodeSequenceSQ (0008,1110) ReferencedStudySequence (0008,1120) ReferencedPatientSequence SQ (0008,1125) ReferencedVisitSequence (0010,0010) PatientName PN 1 (0010,0020) PatientID LO 1 (0010,0021) IssuerOfPatientID LO (0010,0030) PatientBirthDate (0010,0032) PatientBirthTime (0010,0040) PatientSex (0010,0050) PatientInsurancePlanCodeSequence...
  • Page 194 GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL (0038,0011) IssuerOfAdmissionID (0038,0016) RouteOfAdmissionsLO (0038,0020) AdmittingDateDA (0038,0021) AdmittingTime (0038,0050) SpecialNeeds LO (0038,0300) CurrentPatientLocation LO (0038,0400) PatientInstitutionResidence LO (0038,0500) PatientState (0038,4000) VisitComments (0040,0100) ScheduledProcedureStepSequence SQ 1 >(0008,0060) Modality CS 1 >(0032,1070) RequestedContrastAgent...
  • Page 195 GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL (0040,3001) ConfidentialityConstraintOnPatientData Appendix 2 DICOM Value Representations Chapter 8 - - Troubleshooting Page 195...
  • Page 196 GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL Page 196 Section 2.0 - Software Issues...
  • Page 197 GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL Chapter 8 - - Troubleshooting Page 197...
  • Page 198 GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL Page 198 Section 2.0 - Software Issues...
  • Page 199 GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL Chapter 8 - - Troubleshooting Page 199...
  • Page 200 GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL Page 200 Section 2.0 - Software Issues...
  • Page 201 GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL Chapter 8 - - Troubleshooting Page 201...
  • Page 202 GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL Page 202 Section 2.0 - Software Issues...
  • Page 203 GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL 2.2.5 How to check DICOM tags sent by RIS/HIS 1.) Turn off Auto Refresh of worklist in Preferenes->Worklist->Worklist Default Query 2.) Open engineering UI with shortcut ‘shfit+ctrl+F5’, select 2nd icon to open a terminal.
  • Page 204 GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL Switch to terminal window opened previously with shortcut ‘shift+alt+ESC’ and list new patient entiry just queired with command ‘ls’ 10.) Copy this file to /tmp folder with command ‘cp file_name /tmp’...
  • Page 205 GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL Need RIS update to accept * is wildcard character, so the filter Pat_Local_Id=’*’ means patient local ID could be anything. Chapter 8 - - Troubleshooting Page 205...
  • Page 206 GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL Engineering Toolkit Problem: There are no software tools available if the PC hangs. Solution/Workaround: There is a hot-key that will allow you to open an engineering user interface to help diagnose an issue.
  • Page 207 GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL Using VI Editor 2.4.1 Introduction The VI editor is a screen-based text editor used in this product. It’s primarily intended for programmers. The following information is intended to familiarize you with only the most basic VI functions.
  • Page 208 GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL Save Changes The command to save the contents of the editor is <:w>. You can combine the above command with the quit command, or <:wq>. You can specify a different file name to save to by specifying the name after the: w.
  • Page 209 GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL <dw> deletes from current cursor position to the end of the word. <3d> deletes three lines from current cursor position downwards. There is also the <y> command which operates similarly to the d command which take text from the file without deleting the text.
  • Page 210 GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL Item Description Append at the end of the current line. I Insert from the beginning of a line. O (letter <A> oh) Enter insert mode in a new line above the current cursor position.
  • Page 211 GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL Item Description <L> Move the cursor to the first non-white space character on the bottom of the screen. <M> Move the cursor to the first non-whitespace character on the middle of the screen.
  • Page 212 GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL Item Description <~ > Switch the case of the character under the cursor. << > Shift the lines up to where to the left by one shiftwidth. <<<>...
  • Page 213: Wireless Architecture

    GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL Section 3.0 Wireless Architecture 1.) Using scenario for Fix RAD room with monitor and AP. • Detector’s default IP is fixed to 3.1.1.30/24. • AP’s default IP is fixed to 3.1.1.1/24.
  • Page 214: Manual Bad Pixel Calibration

    GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL Section 4.0 Manual Bad Pixel Calibration 1.) Insert the service key, launch the SUIF. 2.) Quit SUIF 3.) Open the raw image in image viewer to mark the bad line or bad pixel Note: To mark bad line, use the line mark tool to completely overlap the bad line.
  • Page 215 GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL 6.) Mark the bad line or bad pixel using annotation tool. 7.) Quit image viewer and click “yes” to save. Chapter 8 - - Troubleshooting Page 215...
  • Page 216 GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL 8.) launch the SUIF. 9.) Click "Manual Bad Pixel Calibration", then click "Start" button. Page 216 Section 4.0 - Manual Bad Pixel Calibration...
  • Page 217 GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL 10.) Click "Continue" button. 11.) Click "Continue" button Chapter 8 - - Troubleshooting Page 217...
  • Page 218 GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL 12.) Click "Continue" button. 13.) Click "Overwrite", then click "Continue" button. Note: Either overwrite or append is ok for first time of MBP, suggest choose Append for second time MBP.
  • Page 219 GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL 4.) Click "Continue" button. 14.) Click "Complete" button. Section 5.0Image Quality Issues Equi-spaced Black Bad Line Issue Although detector calibration/QAP passed, after taking an exposure, there are equi-spaced bad lines at the top of image.
  • Page 220 GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL QAP Pass Issued Image Solutions: 1.) Perform Manual bad pixel calibration. See Section 4.0. 2.) Take exposure again to confirm the bad lines disappear. 3.) Perform QAP again.
  • Page 221 GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL Root Cause: Press exposure switch too quickly after exposure count down clock is displayed in screen during Detector calibration, so detector scan data is lost. Solutions: 1.) Do not press exposure switch immediately after exposure count down clock is displayed in screen.
  • Page 222: Q&A

    GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL Section 6.0 Q&A 1.) Digital Cassette Detector communication lost. Confirm communication with detector. Call service if problem persists. Solution: Goes to “Service ->Diagnostics->Wireless Link Diagnostics”, then follow the recommend procedures in the instruction.
  • Page 223 GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL iii.) Please note that if you choose to Abort, the image you just acquired for the patient will be lost. iv.) If still cannot recover wireless reconnection, and you choose not waiting, you can abort this and go to “Service ->Diagnostics->Wireless Link Diagnostics”, then follow the...
  • Page 224 GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL Solution: Check the exposure techniques is appropriate ii.) Check the maximum exposure timeout is set appropriate iii.) Check the grid of the target system, configure the Gridline remove in Utilities 12.) I do not want to display the patient orientation frame on the image...
  • Page 225 GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL with superuser. ii.) Press keyboard “i” to enable edit function, add a line: route add -host “IP of pacs” gw “ip of gateway”. iii.) Press ESC key and then input :wq to save and exit from configuration file.
  • Page 226 GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL This page is intentionally left blank. Page 226 Section 6.0 - Q&A...
  • Page 227: Chapter 9 Periodic Maintenance

    GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL Chapter 9 - Periodic Maintenance This chapter of the manual provides a recommended schedule for periodic maintenance for the Brivo XR118 System. Section 1.0 General To use this equipment with an optimal situation at any time, the periodic maintenance must be required.
  • Page 228: Preventive Maintenance Action List

    GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL Section 2.0 Preventive Maintenance Action List Operation Frequency Duration Description (Month) (Min) 1.) General A.) Customer Discussion - Discuss the objectives of the PM with the customer: The purpose of planned maintenance is inspection and not repair.
  • Page 229 GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL Operation Frequency Duration Description (Month) (Min) A.) Clean and inspect. B.) Fix hard disk to keep connection tight into workstation. Refer to Portable workstation SSD drive Workstation replacement procedure on page 238.
  • Page 230: Preventive Maintenance

    GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL Section 3.0 Preventive Maintenance Maintenance Is To Be Performed Only By Qualified Trained Personnel Who Are Familiar With The Potential Hazards Associated With This Equipment. Maintenance Activities In The Table Above Are To Be Performed By Authorized Service Personnel Only.
  • Page 231 GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL General 3.1.1 General Cleaning and Painting Keep all surfaces clean. Remove power from the system before beginning work. Check that power is off. 1.) Remove power from the system.
  • Page 232 GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL Digital Box 3.2.1 Software Function Create a new test patient. Take an exposure. Save this image serial. 3.2.2 Hospital Ethernet Push the image serial created above to workstation or printer. Check if it is ok.
  • Page 233: Chapter 10 Renewal Parts

    GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL Chapter 10 - Renewal parts This chapter contains the list of recommended renewal parts for the various models of Brivo XR118 System. Section 1.0 System Components Item Part No.
  • Page 234 GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL 5500786 5500788 5501567 5498484 5501606 5501607 5501661 5501899 5505863 5500786-6 5500786-5 5500786-10 5500786-8 5500786-7 5539067 Page 234 Chapter 10 - - Renewal parts...
  • Page 235 GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL 5505449 5537182 5444248 & 5445907 The cover (FRU part) removal and other renewal parts replacement must be operated by sufficient training personnel. Follow the instructions in Equipment Safety - Electrostatic Discharge (ESD) to prevent static electricity damage to equipment.
  • Page 236 GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL Portable workstation battery replacement procedure Embedded workstation battery information: Rating: 10.8 V Capacity: 95.58Wh Material: Li-ion Type: Workstation battery Tools required: Screwdriver T10 1.) Demount the two screws at the side of the workstation.
  • Page 237 GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL 3.) Disconnect the plug and the socket. 4.) Pull out the battery. 5.) Insert a new battery to replace the old one. The plug end should be close to the socket.
  • Page 238 GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL Portable workstation SSD drive replacement procedure The device or portions of the device may have power present from the battery even when shut down, please demount the battery first before this procedure.
  • Page 239 GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL 4.) Demount the SSD drive from the main board. 5.) Take out the SSD drive and insert a new one. 6.) Secure the SSD drive. 7.) Mount cover with the screws.
  • Page 240 GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL Portable workstation RAM replacement procedure The device or portions of the device may have power present from the battery even when shut down, please demount the battery first before this procedure.
  • Page 241 GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL 4.) Remove the RAM cover. 5.) Take out the RAM from board. Push these two stickers away RAM pops up 6.) Replace a new RAM. Press the RAM down Insert the RAM into slot 7.) Remount the RAM cover.
  • Page 242 GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL Portable workstation main board assembly replacement procedure The device or portions of the device may have power present from the battery even when shut down, please demount the battery first before this procedure.
  • Page 243 GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL 6.) Remove the screws which secure the main board assembly with the front cover. 7.) Take out the main board assembly from the front cover. 8.) Place the new main board assembly into the front cover.
  • Page 244 GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL Wireless AP Replacement Procedure For Wall Mounting 1.) Disconnect the ethernet cable and USB cable with AP. 2.) Demount the AP with wall adapter. 3.) Mount the new AP with wall adapter.
  • Page 245: Digital Detector

    GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL Section 2.0 Digital Detector The device or portions of the device may have power present from the batteries even when shut down, please demount the battery first. Detector Assembly Before the replacement, make sure detector’s malfunction is not caused by collision.
  • Page 246 GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL Detector Battery Locker Assembly Replacement Material/Tools/SW/systems needed: • Phillips screwdriver Note: Below steps show how to replace battery left locker assembly, replacement of the right locker assembly follows the same steps.
  • Page 247 GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL 3.) Remove old locker from detector and put new locker in the slot. Put new locker in the slot After putting the new locker 4.) Push label with linear paper up and put the screw into hole, then tighten the screw with phillips screwdriver (Torque of Screw: 0.2N·M±0.05N·M).
  • Page 248 GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL Put in and tighten the screw Release linear paper and paste the label to cover the screw. 6.) Repeat step 4) and step 5) for the other screw.
  • Page 249: Chapter 11 Master Interconnect System Map

    GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL Chapter 11 - Master Interconnect System Map Section 1.0 Scope This section describes the MIS (Master Interconnection System) map. Section 2.0 Symbols/Conventions The equipment outlines on the map identify where connections are made. These outlines are usually major assemblies. Internal divisions show modules/circuit boards.
  • Page 250: Digital System Mis Map

    12/Aug/2014 Initial relase FIRST MADE FOR: Brivo XR118 APPR Yang Yang 02/Dec/2015 Only support fix RAD upgrade GE HEALTHCARE (TIANJIN) 6118888SCH SHT/CONT ON OTHER COMPANY LIMITED Figure 11-24 Functional Block Diagram Page 250 Section 3.0 - Digital System MIS Map...
  • Page 251: Chapter 12 Set Preferences

    GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL Chapter 12 - Set Preferences Preferences will allow a super-user to customize your system for your facilities needs. It will allow you to predefine your X-ray procedure parameters so that any stored procedure may be retrieved from memory. This will allow you to access technique factors programmed for that type of procedure.
  • Page 252 GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL • Image Processing Change Default Factory Looks for Exams Build Custom Looks Tissue Equalization Overview EMI Reduction Grid Line Reduction Single Energy Grid Line Rate Search Range •...
  • Page 253 GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL Accessing Preferences Preferences are set on the Utilities screen (Figure 12-2), which is accessed by pressing the utilities button on the Worklist. Figure 12-1 Utilities button Note: You must be logged in as an user with the appropriate level of access in order to set preferences.
  • Page 254 GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL General This section provides instructions for setting your system preferences for network and printer connections. The Services Desktop and Shut Down functions are also available from this screen. Refer to...
  • Page 255 GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL Add or Edit Network Host Adding and editing network hosts use very similar process and the same screens as shown in Figure 12-4 Figure 12-5. Table 12-1 Table 12-2 describe the fields in detail.
  • Page 256 GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL Function Description Storage Commitment - Application Entity Type in DICOM AE. Title Storage Commitment - The IP address of the storage database. Network Address Storage Commitment - The port number of the storage database.
  • Page 257 GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL Function Description Perform de- Automatically de-identifies any images that the XR118 identification system sends to this host. (anonymous patient Refer to Chapter 5: Preference Tab for more information images) when sending about de-identification.
  • Page 258 GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL Remove Network Host Follow this process to remove a network host. 1. From the Network Connections screen, select the network host. 2. Click [Remove].  A removal confirm message appears.
  • Page 259 GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL DICOM Printers Follow this process to access the Network Connections screen. 1. On the Worklist screen, click utilities button [  The System - General screen appears. 2. Click the [Edit] button of [Network Connections].
  • Page 260 GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL Add or Edit DICOM Printers Adding and editing printers use very similar process and the same screens shown Figure 12-7, Figure 12-8, Figure 12-9 Figure 12-10. Table 12-3,...
  • Page 261 GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL Function Description Printer Pixel Size Designates the pixel size the printer uses. This is (micron) specified by the printer manufacturer. Configuration A place for you to add notes about the printer or Information configuration.
  • Page 262 GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL Figure 12-8 Add Printer screen - Layouts tab Table 12-4 Add Printer screen - Layouts Tab description Function Description Selections allow the number of images that may be printed on Select the desired a single sheet of film or paper.
  • Page 263 GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL Figure 12-9 Add Printer - Film Sizes tab Table 12-5 Add Printer screen - Film sizes description Function Description Film sizes Selects the sizes of film available for the printer.
  • Page 264 GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL Figure 12-10 Add Printer - Printer Settings tab Table 12-6 Add Printer screen - Printer Settings description Function Description # of Copies Defines how many copies to print.
  • Page 265 GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL Function Description Selects the type of media to print on. Available options are: Media Type • None • Clear film • Blue film Selects the destination. Available options are: Destination •...
  • Page 266 GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL System Shut Down Table 12-7 System Shutdown Function Description Follow this process to perform a Shut Down using the [Shut Down] button. • Press [Shut Down]. A shutdown confirm message appears.
  • Page 267 GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL Detector Management Figure 12-11 System - Detector Management Detector Registration Figure 12-12 System - Detector Management - Detector Registration The associated detector is identified by an asterisk (*).
  • Page 268 GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL Adding A New Detector Identified By Name Figure 12-13 System - Detector Management- Add By Name 1. Press the [Add] button on the Detector Registration screen. 2. Enter the mac address of the detector you wish to register to the system.
  • Page 269 GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL Adding A New Detector Identified By Color and Shape Figure 12-14 System - Detector Management- Add By Color and Shape 1. Press the [Add] button on the Detector Registration screen.
  • Page 270 GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL Removing A Detector A Detector must be “Non-associated” detector, to be removed from the system. If a Detector is “Associated” you must change it to “Non-associated” before you can remove it from the system.
  • Page 271 GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL Click the detector Management icon [ ] to list all the detectors registered in the system. To associate a “Non-associated”: 1. Selects a non-associated detector from the detector list which has been boot 2.
  • Page 272 GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL 4. Upon releasing the associated detector successfully, associated detector will be updated as non-associated.  If the detector release failed, the associated detector will not become non-associated. A popup message will be displayed to indicate user of the release failure.
  • Page 273 GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL Target System Configuration The target system configuration page shall have an editable target system maximum exposure time field. Figure 12-18 Target System Configuration CAUTION: Suggest configuring this value with target system's max exposure time.
  • Page 274 GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL Preferences This section provides instructions for setting your system preferences for the Worklist, Image Management, Viewer, DI, Protocols, Image Processing, Health Page, and Login Management. Worklist Worklist preferences are available from the Utilities screen.
  • Page 275 GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL Default Query The Worklist Default Query controls the amount of HIS/RIS information that appears on the Worklist and allows you to enable and configure the auto-refresh function. 1. Complete or edit the Worklist Default Query screen (Figure 12-20).
  • Page 276 GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL Function Description Auto Refresh Turns Auto Refresh on or off. [On]/[Off] If Auto Refresh is [On], sets how often (in minutes) the Refresh every __ worklist refreshes. The interval may be between 1 and 9999 Minutes minutes.
  • Page 277 GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL Preset Names The Preset Names screen (Figure 12-22) allows you to add, remove, or edit the names that appear on the Add Patient/Patient Information screen’s Operators, Performing Physicians, and Referring Physicians drop-down lists (Figure 12-21).
  • Page 278 GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL Add Preset Names 1. Click [Add] for the appropriate group. 2. Type the name to add. 3. Click [Save]. Edit Preset Names 1. Select the name to change.
  • Page 279 GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL Image Management Image Management preferences allow you to enable and configure the Copy Exam, Auto Tag, Auto Print, Auto Push, and Auto Delete functions. Figure 12-23 Preferences - Image Management...
  • Page 280 GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL Auto Tag (Quality Check) Enabling Auto Tag (or Quality Check) provides an indicator on the Image Viewer screen to mark an image of acceptable quality. Auto Print and Auto Delete can be configured to act upon images that have the Quality Check mark.
  • Page 281 GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL Table 12-9 Auto Print Functions Function Description Lists the printers connected to your system. Refer to Printer Network Connections for information about how to configure the available DICOM printer.
  • Page 282 GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL Auto Send (Auto Push) Auto Send automatically transfers images to another network device when the exam is closed. Auto Send is enabled and configured from the Utilities [ ] –...
  • Page 283 GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL Function Description Allows you to choose which images are sent to each network host. You may choose either Raw or Processed, both, or none. Auto Send Leaving both choices unselected means that no images will be sent to the network host.
  • Page 284 GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL Auto Delete Auto Delete automatically deletes images when the image database does not have enough space. Auto Delete is enabled from the Utilities [ ] - Preferences screen.
  • Page 285 GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL Function Description Delete images until Specifies how many images to delete based on database size. database is __% full. Images must be at Specifies how old (in days) an image must be for it to be least __ Days old deleted.
  • Page 286 GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL Image Processing The Image Processing preferences allow you to view the settings of default factory looks or to create up to five (5) custom looks. Image Processing Preferences also allow you to change which look is the default for the anatomical view.
  • Page 287 GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL 5. When finished, press [Map] to save the changes. • A message appears: “Default look settings for selected Representative Anatomical Views will be mapped to related anatomical views.“...
  • Page 288 GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL 1. From the Worklist screen, click utilities button [ Preferences > Image Processing. 2. Select 3. Click Image Processing [Edit].  The Image Processing Preference Editor (Figure 12-29) appears.
  • Page 289 GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL Table 12-12 Image Processing parameters Parameter Definition Adjusts image contrast. Also known as window width. • Move the slider right for more contrast (towards pure black and Contrast white).
  • Page 290 GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL Parameter Definition If grid status is <Out>, Grid Line Removal algorithm will be not used. If grid status is <In> and at the same time Grid Line Grid Status Reduction selection in Utility is On, Grid line reduction algorithm will be used.
  • Page 291 GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL Change Image Processing from the Viewer Changing the image processing from the Image Viewer screen applies the process changes to a specific image, but also allows mapping the process change to multiple.
  • Page 292 In combination with the wide dynamic range of the GE Revolution? digital detector, TE allows display of more information collected in a single shot, reducing re-takes and increasing throughput.
  • Page 293 GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL TE parameters TE uses the information in the image to improve visualization. Two user-defined parameters; AREA and STRENGTH; control the extent and amount of TE that is applied to the image. There is an AREA and STRENGTH parameter for over- penetrated regions and an AREA and STRENGTH parameter for under-penetrated regions.
  • Page 294 GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL Figure 12-33 Varied TE settings Note: Unlike Tissue Equalization, which controls contrast within under-penetrated (thick) or over-penetrated (thin) areas, Tissue Contrast controls the general contrast between the thick and thin anatomy. For example, Tissue Contrast can be used in combination with Tissue Equalization to define a Chest PA look that is equalized in the lung fields, but with light spine/ribs.
  • Page 295 GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL EMI Reduction Image artifacts can be caused when the Digital Detector acts like an antenna, picking up electromagnetic signals emitted by external equipment in the area. Artifacts are periodic lines or bands in the image. EMI artifacts can be...
  • Page 296 GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL Note: If the image is not good due to the GLR setting’s non-consistent with target sys- tem, please re-configure GLR and reprocess the image. Note: Make sure the detector is fully covered by the grid if grid configured.
  • Page 297 GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL Target System Grid Type Recommend Grid Type for GLR 64 LP/CM 36LP/CM 66 LP/CM 36LP/CM 68 LP/CM 70LP/CM 70 LP/CM 70LP/CM 72 LP/CM 70LP/CM, 78LP/CM 74 LP/CM 70LP/CM, 78LP/CM...
  • Page 298 GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL DI (Deviation Index) Figure 12-36 Preferences - DI screen The Deviation Index screen allows you to specify the DI Control Limits, control if or how the DI is displayed on the Image Viewer screen, change the lower and upper limits for anatomical views.
  • Page 299 GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL DI Control Limits DI (Deviation Index) estimates the deviation of an actual detector exposure from an ideal detector exposure. Figure 12-37 Preferences - DI Control Limits The DI Control Limits screen lets you customize your DI Control Limits, as well as the DI Control Range Suggestions.
  • Page 300 GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL DI-Preferences Follow this process to change the DI settings. 1. From the Worklist screen, press utilities button [ 2. Select Preferences > DI. 3. Press DI-Preferences [Edit].  The Deviation Index screen (Figure 12-38) appears.
  • Page 301 GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL Change the DI Display 1. On the Deviation Index screen, click the DI Display drop-down list to open 2. Select the display option.  Available options are: In this example, there is no DEI –...
  • Page 302 GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL Enable or Disable Technical Mode Technical Mode is a special setting that configures the system for image quality testing and detector calibration. When Technical Mode is enabled, DI is calculated using the central area of the image (512 pixels x 512 pixels) regardless of the imaged anatomy.
  • Page 303 GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL Viewer Figure 12-40 Preferences - Viewer screen Chapter 12 - - Set Preferences Page 303...
  • Page 304 GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL Pre-set Annotations Viewer Preferences allow you to add, edit, or remove pre-set annotations. Follow this process to configure pre-set annotations. 1. From the Worklist screen, click utilities button [ Preferences >...
  • Page 305 GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL Add Pre-set Annotation 1. From the Pre-set Annotations screen, click [Add]. 2. Enter the text of the annotation. 3. Click [Save]. Edit Pre-set Annotation 1. From the Pre-set Annotations screen, select the annotation.
  • Page 306 GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL Remove Pre-set Annotation 1. From the Pre-set Annotations screen, select the annotation. 2. Click [Remove].  A removal confirm message appears. 3. Click [OK]. Print/Push Font Size Adjustment 1.
  • Page 307 GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL 3. Click Custom Annotations [Edit].  The Custom Annotations editing screen (Figure 12-43) appears. Figure 12-43 Custom Annotations 4. Select (check) the annotations you want to appear. 5. Adjust the Font Size, if necessary.
  • Page 308 GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL Protocols Protocols preferences allow you to create backup copies of the protocol database, retrieve saved backups, and create new protocols. Figure 12-44 Preferences - Protocols screen Page 308...
  • Page 309 GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL Backup Protocol Database to CD/DVD The Backup function allows you to save the entire protocols database (parameters) to a CD/DVD. This is important when editing protocols; backup ensures that the current database stays intact. Backup is also important in case of system failure and all protocol information is lost.
  • Page 310 GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL Retrieve Protocol Database from CD/DVD The Retrieve function allows you to re-cover a protocol database that was saved to a CD/DVD. Note: When retrieving, the procedures saved on the disk will overwrite all of the procedures on the system.
  • Page 311 GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL Edit Protocol Database The Protocol Editor allows you to create custom acquisition protocols. It allows the creation of anatomical categories, exam sets, and individual views for exams. It allows the edit and removal of categories, exams, and views.
  • Page 312 GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL Add or Edit Category Note: The screens and process to edit a category are the same as for adding a category. From the Exam Menu: 1. Click add button [ ] or edit button [ ] of Category column.
  • Page 313 GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL Add or Edit Exam Note: The screens and process to edit an exam are the same as for adding a exam. From the Exam Menu: 1. Select the category for the new exam.
  • Page 314 GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL Add or Edit View Note: The screens and process to edit a view are the same as for adding a view. Note: This process has multiple screens. From the Exam Menu: 1.
  • Page 315 GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL Laterality 7. Select the  Options are: – Paired (exam has R or L, as with extremities). This selection allows the Patient Side control to become active on the Image Acquisition screen.
  • Page 316 GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL Protocol Editor Copy Functions Follow this process to access the Protocol Database Editor. 1. From the Worklist, select utilities button [ Preferences > Protocols 2. Select 3. Click the [Edit] button of Protocol Database.
  • Page 317 GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL Copy Exam 1. Select the Exam to copy. 2. Press copy button [ ] of exam. 3. Enter the new name to use or retype the existing name.
  • Page 318 GE H EALTHCARE 5461425-8EN, R XR118 S IRECTION EVISION RIVO ERVICE ANUAL Copy View 1. Select the View to copy. 2. Press copy button [ ] of View. View Name 3. Enter the to use or retype the existing name.

Table of Contents