3.3.4 Acquire a scout

Use this procedure to acquire a scout from which you can prescribe the scan locations.
1. Confirm that the patient orientation matches the patient orientation in the
area of the
Click on the patient image to change to a head-first or feet-first orientation.
Click on the table or above the patient body to rotate the body in 90° increments.
Scout collection on the Scan Settings window, confirm the Start and End locations,
2. From the
kV , mA , and Auto Voice settings (if applicable).
To change a parameter, open the desired settings collection and click in the field to enter
a new value. Plus (+) can be used for S values and Minus (-) for I values when using the
ten-digit key pad.
Technique factors are set low (120 kV and a low mA are common), since these scans
are normally used only for planning purposes.
Scout Plane to designate the type of scout to acquire.
3. Set the
0° = 12 o'clock
90° = 3 o'clock
180° = 6 o'clock
270° = 9 o'clock
Window Width and Window Length fields, type the desired values.
4. In the
256
Illustration 9: Scout collection
Scan Settings window.
Revolution CT User Manual
Direction 5480385-1EN, Revision 1
Patient Position
3 Exam Workflow

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Revolution CT and is the answer not in the manual?

Questions and answers

Greg Lesco
February 27, 2025

need a part number for the Integrated Cardiac Patient Monitor, patient grounding strap + cable

Yaya
March 14, 2025

How do I clear Ge Revolution collision sensor

1 comments:
Mr. Anderson
March 14, 2025

To clear the collision sensor on a GE Revolution CT, remove any obstructions detected by the sensors located beneath the front and back of the table. Once the area is clear, you can continue positioning the table.

This answer is automatically generated

Table of Contents

Save PDF