Select a recon type that is appropriate for coronary imaging.
Poor contrast opacification of the coronary vessels (low or non-uniform HU enhancement
from poor or missed bolus, venous contamination, et cetera) can result in sub-optimal
SnapShot Freeze motion correction.
Streak artifacts from surgical clips, pacemaker lead wires, etc., in close proximity to the
coronaries, can result in sub-optimal SnapShot Freeze motion correction.
Utilize targeted reconstruction to maximize pixel resolution; however the prescribed DFOV
should allow for at least 2 cm of tissue between the edge of the DFOV and the heart. Do not
"over-zoom".
As SnapShot Freeze targets coronary motion, for an acquisition series with a scan range
significantly longer than the heart itself, select only the axial scan location(s) containing the
heart to streamline subsequent SSF motion correction processing.
Millisecond-based phase prescription and the ECG Editor are fully compatible with
SnapShot Freeze. Utilize as needed.
Prescribe SnapShot Freeze for a Secondary Recon
Use this procedure to prescribe SnapShot Freeze for a Secondary Recon to reduce motion in
the coronary vessels. SnapShot Freeze is only available for single phase and center phase
recon options.
NOTE:
SnapShot Freeze is not available for a Primary Recon.
1. From the image monitor, in the
appropriate Secondary Recon task (prescribed as part of the protocol) or add a Secondary
Recon task using [Add a Recon].
2. Click on an existing Secondary Recon task or add a new one to open the editing panel to
the right of the task list.
Adjust Thickness and Images per rotation as needed.
ASiR-V can be turned "On" by clicking the selection, choosing an ASiR-V level in %, and
then clicking [Apply].
Adjust any other parameters as needed.
Anatomy Selection collection to adjust the recon Start Location, End Location,
3. Open the
DFOV, A/P Center offset, and R/L Center offset as needed.
The DFOV has to be at least 15 cm.
The prescribed DFOV should allow for at least 2 cm of tissue between the edge of the
DFOV and the heart.
ECG and Gating collection.
4. Open the
Select SnapShot Freeze to "On".
Select Phase Type as needed.
368
Reconstruction and Imaging Processing area, select the
Revolution CT User Manual
Direction 5480385-1EN, Revision 1
7 SnapShot Freeze