When the button is in this state:
Auto-Run
Run
3.2 Reformat planning
3.2.1 Automatic reformat
For an automatic reformat, no images are acquired before planning the reformat.
1. From the Reconstruction and Image Processing Task List, click the down arrow on the
series you want to reformat.
2. Select the protocol you want to use from the
3. Check the
will be generated automatically.
3.2.2 Semi-automatic reformat
1. From the Reconstruction and Image Processing Task List, click the down arrow on the
series you want to reformat.
2. Select the protocol you want to use from the
the protocol. Auto Run is not selected.
3. After the source images have been reconstructed, select the reformat task.
428
The reformat is in this state:
You are prescribing the reformat task. If the button is grayed out, select the
The recon data is available, but the re‐
format has not been run.
Auto Run box in the lower right of the window to enable auto run. DMPR images
Direction 5480385-1EN, Revision 1
Clicking the button does this:
appropriate protocol from the Protocol
section.
Click the button to autorun the refor‐
mat after recon data has been ac‐
quired.
Click the button to run the reformat.
This button is grayed out if you have
not selected a protocol or defined a
batch reformat manually, or an Auto
Batch has been run.
Reformat Settings panel.
Reformat Settings panel. Adjust the settings for
3 Direct Multi Planar Reformat (DMPR)
Revolution CT User Manual
Need help?
Do you have a question about the Revolution CT and is the answer not in the manual?
Questions and answers
need a part number for the Integrated Cardiac Patient Monitor, patient grounding strap + cable
How do I clear Ge Revolution collision sensor
To clear the collision sensor on a GE Revolution CT, remove any obstructions detected by the sensors located beneath the front and back of the table. Once the area is clear, you can continue positioning the table.
This answer is automatically generated