DFOV
•
R/L Center
•
A/P Center
Once the recon and multi-planar reformat batch images are created, the Reconstruction and
Image Processing Task List allows you to select and view image sets.
Considerations
•
Each group must be contiguous to be able to be combined with the prior group in a DMPR
session. DMPR can be prescribed with primary and secondary recons. DMPR reformats are
limited to 2,000 images. Additional reformat protocols can be added from the Volume Viewer
Reformat application which can be accessed with the File Manager.
•
Any scans acquired after scanning has completed on the original scan group using Add
Group are not added to the DMPR session. Remember to include all of the desired
coverage area in the original scan prescription.
•
If the patient orientation is prescribed as Decubitus (right or left), you will observe that the
Paging slider for the sagittal image will scroll images in DMPR coronal viewport and vice
versa. This is because in Decubitus orientation, patient's sagittals and coronals are
switched. Hence, the DMPR sagittal viewport contains the coronals from patient's reference
axis, and the DMPR coronal viewport contains the sagittals from the patient's reference axis.
•
Make sure the interval between groups is appropriate to support DMPR in secondary recons
2-10 if the slice thickness prescribed for the groups in primary recon 1 is different.
3.1 Reformat Tools panel
To display the Reformat Tools panel, select or create a reformat task from the Reconstruction
and Image Processing task list.
420
Revolution CT User Manual
Direction 5480385-1EN, Revision 1
3 Direct Multi Planar Reformat (DMPR)
Need help?
Do you have a question about the Revolution CT and is the answer not in the manual?
Questions and answers
need a part number for the Integrated Cardiac Patient Monitor, patient grounding strap + cable
How do I clear Ge Revolution collision sensor
To clear the collision sensor on a GE Revolution CT, remove any obstructions detected by the sensors located beneath the front and back of the table. Once the area is clear, you can continue positioning the table.
This answer is automatically generated